TRIGger:WIDTh:MIN <MinDuration>
Sets the lower limit for the pulse width if TRIGger:WIDTh:RANGe is set to WITHin or
OUTSide.
Parameters:
<MinDuration> Range: 800E-12 to 10000
Increment: 100E-9
*RST: 5E-9
Default unit: s
14.3.5.5 Video/TV Trigger
The standards PAL, PAL-M, NTSC and SECAM are available in the instrument firm-
ware. All other standards and custom signals require the advanced trigger option
(R&S RTH-K19).
TRIGger:TV:STANdard...................................................................................................267
TRIGger:TV:POLarity..................................................................................................... 268
TRIGger:TV:MODE........................................................................................................ 268
TRIGger:TV:LINE...........................................................................................................269
TRIGger:TV:LFIeld.........................................................................................................269
TRIGger:TV:CUSTom:STYPe......................................................................................... 269
TRIGger:TV:CUSTom:LDURation....................................................................................270
TRIGger:TV:CUSTom:SDURation................................................................................... 270
TRIGger:TV:CUSTom:SCANmode...................................................................................270
TRIGger:TV:STANdard <Standard>
Selects the TV standard or CUSTom for user-defined signals.
Waveform Setup