User Manual
Remote Control Commands
R&S
®
Scope Rider RTH
270User Manual 1326.1578.02 ─ 08
Parameters:
<SyncPulseType> BILevel | TRILevel
*RST: BILevel
TRIGger:TV:CUSTom:LDURation <LineDuration>
Sets the duration of a single video line, the time between two successive sync pulses.
Parameters:
<LineDuration> Range: 5E-6 to 200E-6
Increment: 100E-9
*RST: 64E-6
Default unit: s
TRIGger:TV:CUSTom:SDURation
<SyncPulseDuration>
Sets the width of the sync pulse.
Parameters:
<SyncPulseDuration> Range: 100E-9 to 100E-6
Increment: 100E-9
*RST: 4.7E-6
Default unit: s
TRIGger:TV:CUSTom:SCANmode <ScanMode>
Sets the scanning system.
For details, see "Scan" on page 57.
Parameters:
<ScanMode> INTerlaced | PROGressive | SEGMented
*RST: INTerlaced
14.3.5.6 External Trigger (R&S RTH1002)
TRIGger:EXTernal:LEVel................................................................................................270
TRIGger:EXTernal:SLOPe.............................................................................................. 271
TRIGger:EXTernal:LEVel <Level>
Sets the trigger voltage level.
Parameters:
<Level> Range: -10 to 10
Increment: 1E-3
*RST: 0
Default unit: V
Waveform Setup