User Manual
Remote Control Commands
R&S
®
Scope Rider RTH
353User Manual 1326.1578.02 ─ 08
Parameters:
<ReferenceValue> Range: +/- (1.1 * measurement range)
*RST: 0
Default unit: V (VOLTage) | A (CURRent)
METer<m>:SENSe:NULL:VALU <ReferenceValue>
Sets the reference value for the currently active measurement type.
Relative measurements are not available for diode, continuity, and frequency measure-
ments.
Suffix:
<m>
.
1..4
R&S RTH1002: always 1, omit the suffix
R&S RTH1004: Selects the voltmeter.
Parameters:
<ReferenceValue> Range: +/- (1.1 * measurement range)
Default unit: Depends on the measurement type
14.8.4 Measurement Control
METer<m>:SENSe:FUNCtion......................................................................................... 353
METer<m>:CONFigure:VALue?...................................................................................... 354
METer<m>:TRIGger:MODE............................................................................................ 354
METer<m>:ABORt......................................................................................................... 354
METer<m>:SENSe:FUNCtion <MeasType>
Sets the measurement type for the selected meter. All measurement attributes of the
previous function (range, resolution, etc.) are remembered. If you return to the previous
function, the measurement attributes are restored.
Changing the measurement type disables scaling, limit testing, histogram, statistics,
and trend chart data collection: CALC:<function>:STAT is set OFF.
Suffix:
<m>
.
1..4
Selects the voltmeter.
Setting parameters:
<MeasType> 'VOLTage:AC | VOLTage[:DC] | VOLTage:AD'
String parameter
*RST: VOLTage [:DC]
Example:
MET2:SENS:FUNC "VOLT:AC"
MET2:CONF:VAL?
<--"VOLT:AC 1"
Sets the second voltmeter to AC voltage measurement.
Usage: Setting only
Voltmeter (R&S RTH1004)