User Manual
Remote Control Commands
R&S
®
Scope Rider RTH
358User Manual 1326.1578.02 ─ 08
Usage: Query only
METer<m>:CALCulate:AVERage:MAXimum?
Returns the maximum value of the measurement series.
Suffix:
<m>
.
1..4
R&S RTH1002: always 1, omit the suffix
R&S RTH1004: Selects the voltmeter.
Return values:
<Maximum> Range: -100E+24 to 100E+24
Increment: 1E-12
*RST: 0
Default unit: V
Usage: Query only
METer<m>:CALCulate:AVERage:CLEar
Deletes all statistical values.
Statistics are also deleted if:
●
The measurement function changes (METer:SENSe:FUNCtion).
●
*RST
●
SYSTem:PRESet
Suffix:
<m>
.
1..4
R&S RTH1002: always 1, omit the suffix
R&S RTH1004: Selects the voltmeter.
Usage: Event
14.9 Counter Mode (R&S RTH-K33)
COUNter<m>:ABORt..................................................................................................... 359
COUNter<m>:CALCulate:AVERage:ALL?........................................................................ 359
COUNter<m>:CALCulate:AVERage:AVERage?................................................................359
COUNter<m>:CALCulate:AVERage:CLEar.......................................................................359
COUNter<m>:CALCulate:AVERage:COUNt:CURRent?.....................................................360
COUNter<m>:CALCulate:AVERage:MAXimum?............................................................... 360
COUNter<m>:CALCulate:AVERage:MINimum?................................................................ 360
COUNter<m>:CONFigure:FREQuency.............................................................................361
COUNter<m>:CONFigure:VALue?...................................................................................361
COUNter<m>:FETCh?................................................................................................... 361
COUNter<m>:INITiate.................................................................................................... 362
COUNter<m>:MEASure:FREQuency?............................................................................. 362
Counter Mode (R&S RTH-K33)