Remote Control Commands
R&S
®
Scope Rider RTH
401User Manual 1326.1578.02 ─ 08
BUS:CAN:TECHnology <ThresholdType>
Parameters:
<ThresholdType> TTL | ECL | CMOS | CAN | GND | LIN7vsupply | LIN12vsupply |
LIN18vsupply | USER
TTL
1.4 V
ECL
-1.3 V
CMOS
2.5 V
GND
0 V (for CAN channels, requires option R&S RTH-K3)
CAN
2 V (for CAN channels, requires option R&S RTH-K3)
LIN7vsupply | LIN12vsupply | LIN18vsupply
7 V / 12 V / 18 V (for LIN channels, requires option R&S RTH-
K3)
USER
Set the value with CHANnel<m>:THReshold:USER.
*RST:
CAN
BUS:CAN:DATA:THReshold <ThresholdValue>
Sets the threshold value for digitization of signals manually. If the signal value is higher
than the threshold, the signal state is high. Otherwise, the signal state is considered
low.
This value is only considered for BUS:CAN:TECHnology USER
Parameters:
<ThresholdValue> Range: -400 to 400
Increment: 1E-3
*RST: 1.4
Default unit: V
BUS:SETReflevels
Sets the appropriate threshold or threshold for the selected serial protocol.
Usage:
Event
BUS:CAN:SAMPlepoint <SamplePoint>
The sample point divides the nominal bit period into two distinct time segments. The
length of the time segments is defined in time quanta according to network and node
conditions during CAN development.
Protocol Analysis