TRIGger:CAN:DATA <Pattern>
Specifies the data pattern to be found, in binary or hex format. Enter the pattern in
MSB first bit order.
Parameters:
<Pattern> String that contains the pattern in binary format. The parameter
accepts the bit value X (don't care).
TRIGger:CAN:DCONdition <Relation>
Defines how the specified data pattern is compared with the acquired signal. The
instrument triggers if the acquired address is equal or unequal the defined pattern.
Parameters:
<Relation> EQUal | NEQual
*RST: EQUal
TRIGger:CAN:FORMerror
<FormError>
A form error occurs when a fixed-form bit field contains one or more illegal bits.
Parameters:
<FormError> ON | OFF
*RST: ON
TRIGger:CAN:FTYPe <FrameType>
CAN has several frame types which can be used as trigger condition.
For data and remote frames, the identifier format has to be set with TRIGger:CAN:
ITYPe.
Protocol Analysis