User Manual
Remote Control Commands
R&S
®
Scope Rider RTH
413User Manual 1326.1578.02 ─ 08
Parameters:
<ThresholdValue> Range: -400 to 400
Increment: 1E-3
*RST: 1.4
Default unit: V
BUS:LIN:TECHnology <ThresholdType>
Sets the threshold value for digitization of signals according to the specified technol-
ogy. If the signal value is higher than the threshold, the signal state is high. Otherwise,
the signal state is considered low.
To set a user-defined threshold, select USER and define the value using BUS:LIN:
DATA:THReshold on page 412.
Parameters:
<ThresholdType> TTL | ECL | CMOS | CAN | GND | LIN7vsupply | LIN12vsupply |
LIN18vsupply | USER
CMOS
2.5 V
LIN7vsupply | LIN12vsupply | LIN18vsupply
7 V / 12 V / 18 V
USER
Set the value with BUS:LIN:DATA:THReshold.
*RST: TTL
BUS:SETReflevels
Sets the appropriate threshold or threshold for the selected serial protocol.
Usage:
Event
14.11.6.2 LIN Trigger
TRIGger:LIN:CHKSerror................................................................................................. 413
TRIGger:LIN:DATA........................................................................................................ 414
TRIGger:LIN:DCONdition................................................................................................414
TRIGger:LIN:ICONdition................................................................................................. 414
TRIGger:LIN:IDENtifier................................................................................................... 414
TRIGger:LIN:IPERror..................................................................................................... 415
TRIGger:LIN:SYERror.................................................................................................... 415
TRIGger:LIN:TYPE.........................................................................................................415
TRIGger:LIN:CHKSerror <ChecksumError>
Triggers on a checksum error if TRIGger:LIN:TYPE is set to ERRC
Protocol Analysis