User Manual

Waveform Setup
R&S
®
Scope Rider RTH
41User Manual 1326.1578.02 ─ 08
RANGE keys
The vertical RANGE keys set the vertical scale (vertical sensitivity) of the selected
waveform.
In FFT mode, the RANGE keys set the scale for the amplitude range (y-axis) in the
spectrum display.
In "Counter" mode, the RANGE keys set the measurement range.
Remote command:
CHANnel<m>:SCALe on page 253
CHANnel<m>:RANGe on page 253
FFT mode:
SPECtrum:FREQuency:MAGNitude:SCALe on page 305
Spectrum mode:
SPECtrum:FREQuency:SCALe on page 310
Counter mode:
COUNter<m>:SENSe:RANGe on page 363
POS keys
Move the selected signal up or down in the diagram. The position is a graphical setting
given in divisions, while the offset sets a voltage.
You can also drag the channel marker on the screen.
Remote command:
CHANnel<m>:POSition on page 253
Spectrum mode:
SPECtrum:FREQuency:POSition on page 310
2.2.2 Effect of the Bandwidth Filter
Low-pass filters reduce the speed of the signal inside the instrument and cause a delay
of the signal on the screen. The delay time depends on the selected filter.
The following table lists the approximate delay of the signal that is caused by various
filters.
Table 2-1: Approximate signal delay dependent on the bandwidth filter
Filter Approx. delay
Filter Approx. delay
200 MHz 30.2 ns
500 kHz 9.07 μs
100 MHz 30.7 ns
200 kHz 22.13 μs
50 MHz 138.5 ns
100 kHz 43.87 μs
20 MHz 145 ns
50 kHz 87.47 μs
10 MHz 166.5 ns
20 kHz 218 μs
5 MHz 193 ns
10 kHz 434.7 μs
2 MHz 270.5 ns
5 kHz 869.3 μs