User Manual
Waveform Setup
R&S
®
Scope Rider RTH
53User Manual 1326.1578.02 ─ 08
±Δ
1 2
±Δ
Figure 2-6: Pulse width is equal or unequal to a given width, with optional tolerance
1 = Equal, pulse > width - Δ AND pulse < width + Δ
2 = Unequal, pulse < width - Δ OR pulse > width + Δ
Remote command:
TRIGger:WIDTh:RANGe on page 265
Width
Sets the width for comparisons equal, unequal, shorter, and longer.
Remote command:
TRIGger:WIDTh:WIDTh on page 266
±Tolerance
Sets a range Δt to the specified Width if the comparison range is equal or unequal. To
trigger on an exact pulse width, set the tolerance to 0.
Remote command:
TRIGger:WIDTh:DELTa on page 266
Min Width / Max Width
Set the lower and upper time limits defining the time range if "Inside" or "Outside" is set
for comparison.
Remote command:
TRIGger:WIDTh:MIN on page 267
TRIGger:WIDTh:MAX on page 266
2.5.5 Video/TV Trigger
The TV or video trigger is used to analyze analog baseband video signals. You can
trigger on baseband video signals from standard definition and high definition stand-
ards, and also on user defined signals.
The instrument triggers on the line start - the horizontal sync pulse. You can trigger on
all lines, or specify a line number. You can also trigger on the field or frame start.
2.5.5.1 Standard TV Trigger Settings
Access: SETUP key > "Trigger Type" = "Video/TV"
Trigger