User Manual
Waveform Setup
R&S
®
Scope Rider RTH
60User Manual 1326.1578.02 ─ 08
"Unequal"
If the pattern is true for a time shorter than "Pattern Width" - Δt OR
longer than "Pattern Width" + Δt, the trigger occurs.
"Inside"
If the pattern is true for a time between "Min Pattern Width" and "Max
Pattern Width", the trigger occurs.
"Outside"
If the pattern is true for a time shorter than "Min Pattern Width" OR
longer than "Max Pattern Width", the trigger occurs.
Remote command:
TRIGger:PATTern:WIDTh:RANGe on page 272
TRIGger:PATTern:TIMeout[:TIME] on page 272
TRIGger:PATTern:WIDTh[:WIDTh] on page 272
TRIGger:PATTern:WIDTh:DELTa on page 273
TRIGger:PATTern:WIDTh:MINWidth on page 273
TRIGger:PATTern:WIDTh:MAXWidth on page 273
2.5.8 State Trigger (R&S RTH-K19)
The state trigger verifies if the channel states match the defined pattern at the clock
edge. The trigger occurs if the logical combination of the input channels is true at the
crossing point of the selected clock edge and the trigger level.
► SETUP key > "Trigger Type" = "State"
Description of settings
Figure 2-12: State trigger
Clock Source
Selects the input channel of the clock signal.
Remote command:
TRIGger:STATe:CSOurce[:VALue] on page 274
Clock Slope
Sets the edge of the clock at which the instrument checks the signal states: at the ris-
ing edge, the falling edge, or at both edges.
Trigger