RT-Zxx_High_bandwidth_dat-sw_en_3607-3868-22_v1700_Cover.indd 1 Data Sheet | 17.00 Test & Measurement R&S®RT-Zxx High-Bandwidth Probes Specifications 09.11.
Version 17.00, November 2016 CONTENTS Definitions.............................................................................................................................................................................................. 3 Probe/oscilloscope chart ........................................................................................................................................................................ 4 R&S®RT-ZZ80 transmission line probe .....................................
Version 17.00, November 2016 Definitions General Product data applies under the following conditions: Three hours storage at ambient temperature followed by 30 minutes warm-up operation Specified environmental conditions met Recommended calibration interval adhered to Specifications with limits Represent warranted product performance by means of a range of values for the specified parameter.
Version 17.
Modular probes RT-ZM15 RT-ZM30 RT-ZM60 RT-ZM90 ● ○ ● ○ ○ ○ ● ○ ○ ○ ● ○ ○ ○ ● ○ ○ ○ ● ○ ○ ○ ○ ● ○ ○ Page 4 GHz 2 GHz 1 GHz 600 MHz 500 MHz 350 MHz RT-ZA9 RTE/RTO 200 MHz 1 GHz 500 MHz 350 MHz RTM 200 MHz Probe: R&S® HMO2000 Base unit: R&S® HMO1000 Version 17.
Version 17.00, November 2016 R&S®RT-ZZ80 transmission line probe All parameters are valid for the probe only when connected to a host instrument with an input impedance of 50 Ω. See table on page 4 and Rohde & Schwarz oscilloscope operating manual for more details.
Version 17.
Version 17.00, November 2016 R&S®RT-ZS10L active probe All parameters are valid when the probe is connected to an appropriate Rohde & Schwarz oscilloscope with an input impedance of 50 Ω. See table on page 4 and Rohde & Schwarz oscilloscope operating manual for more details.
Version 17.00, November 2016 R&S®RT-ZS10L maximum rated sine-wave peak voltage versus frequency.
Version 17.
Version 17.00, November 2016 R&S®RT-ZS10/-ZS10E/-ZS20/-ZS30 active probes All parameters are valid for the probe only when connected to a host instrument with an input impedance of 50 Ω. See table on page 4 and Rohde & Schwarz oscilloscope operating manual for more details.
Version 17.
Version 17.00, November 2016 Sine-wave peak voltage [V] 35 30 25 20 15 10 5 0 1E6 10E6 100E6 1E9 10E9 Frequency [Hz] Maximum nondestructive sine-wave peak voltage versus frequency.
Version 17.00, November 2016 R&S®ProbeMeter Specifications for measurement error apply only when offset compensation setting is 0 V. Specifications for input impedance, dynamic range and maximum nondestructive input voltage apply. The R&S®RT-ZS10E probe is not equipped with an R&S®ProbeMeter. Measurement error +15 °C to +35 °C 0 °C to +50 °C ±0.1 % of reading ± 750 µV ±0.2 % of reading ± 1.
Version 17.00, November 2016 EMC Calibration interval Safety Mechanical data Dimensions Weight Probe interface Connector in line with EMC Directive 2004/108/EC, IEC/EN 61326-1 (table 2), IEC/EN 61326-2-1, CISPR 11/EN 55011(class B) 2 years in line with IEC/EN 61010-1 probe head (W × H × L) cable length overall length probe only approx. 12 mm × 7.5 mm × 68 mm (0.47 in × 0.3 in × 2.68 in) approx. 1.1 m (43 in) approx. 1.3 m (51 in) approx. 90 g (0.
Version 17.00, November 2016 R&S®RT-ZS60 active probe All parameters are valid for the probe only when connected to a host instrument with an input impedance of 50 Ω. See table on page 4 and Rohde & Schwarz oscilloscope operating manual for more details.
Version 17.00, November 2016 Dynamic range DC Offset compensation range Offset compensation error not when offset compensation setting = 0 V AC with zero or compensated DC component Total harmonic distortion 16 V (Vpp) sine-wave input Noise voltage referenced to probe input Maximum nondestructive input voltage DC peak voltage AC peak voltage derated, see figure on page 18 ESD tolerance human body model ±8 V (+ offset compensation setting) ±10 V ±0.2 % of setting ± 2 mV (meas.) 16 V (Vpp) –70 dB (meas.
35 10 M 30 1M |Input impedance| [W] Sine-wave peak voltage [V] Version 17.00, November 2016 25 20 15 10 10E6 100E6 1E9 10E9 Frequency [Hz] Maximum nondestructive sine-wave peak voltage versus frequency. 18 10 k 1k 100 5 0 1E6 100 k Rohde & Schwarz R&S®RT-Zxx High-Bandwidth Probes 10 100E0 100 1E3 10E3 100E3 1E6 10E6 100E6 Frequency [Hz] Input impedance versus frequency.
Version 17.00, November 2016 R&S®RT-ZD10/-ZD20/-ZD30 differential probes All parameters are valid for the probe only when connected to a host instrument with an input impedance of 50 Ω. See table on page 4 and Rohde & Schwarz oscilloscope operating manual for more details.
Version 17.00, November 2016 DC characteristics Attenuation Attenuation error Temperature drift, attenuation Zero error after applying digital correction factors Temperature drift, zero error Dynamic range Differential input Offset compensation range Offset compensation error between signal sockets Operating voltage window CM offset compensation range 3 CM offset compensation error 10:1 ±0.
Version 17.00, November 2016 Sine-wave peak voltage [V] 35 30 25 20 15 10 5 0 1E6 10E6 100E6 1E9 10E9 Frequency [Hz] Maximum nondestructive sine-wave peak voltage versus frequency. Rise time versus input voltage (meas.). R&S®ProbeMeter Specifications for measurement error apply only when offset compensation setting is 0 V. Specifications for input impedance, dynamic range and maximum nondestructive input voltage apply.
Version 17.00, November 2016 R&S®RT-ZA15 external attenuator All parameters are valid for the R&S®RT-ZA15 external attenuator when connected to an R&S®RT-ZD10/-ZD20/-ZD30 differential probe.
Version 17.00, November 2016 Maximum rated input voltage DC voltage AC voltage Transient peak voltage ESD tolerance each signal socket to ground each signal socket to ground, derated, see figure each signal socket to ground human body model, each signal socket to ground ±60 V 30 V (RMS) ±42 V 8 kV (meas.) Maximum rated sine-wave peak voltage versus frequency. General data See page 14.
Version 17.00, November 2016 R&S®RT-ZD40 differential probe All parameters are valid for the probe only when connected to a host instrument with an input impedance of 50 Ω. See table on page 4 and Rohde & Schwarz oscilloscope operating manual for more details.
Version 17.00, November 2016 DC characteristics Attenuation Attenuation error Temperature drift, attenuation Zero error Temperature drift, zero error Dynamic range Differential input Offset compensation range Offset compensation error Operating voltage window CM offset compensation range 4 CM offset compensation error 10:1 after applying digital correction factors 0 °C to +50 °C ±0.
Version 17.00, November 2016 R&S®ProbeMeter See page 21. General data See page 14.
Version 17.00, November 2016 R&S®RT-ZM15/30/60/90 modular probes All parameters are valid for the probe only when connected to a host instrument with an input impedance of 50 Ω. See table on page 4 and Rohde & Schwarz oscilloscope operating manual for more details. R&S®RT-ZM15/30/60/90 probe amplifiers The R&S®RT-ZMxx probe amplifiers support four different measurement modes for differential mode (DM), common mode (CM) and single-ended measurements on the positive (P) and negative (N) signal terminal.
Version 17.00, November 2016 Input impedance DC input resistance DC characteristics Attenuation Attenuation error Temperature drift, attenuation Zero error Temperature drift, zero error Dynamic range Input voltage range Operating voltage window Offset compensation range Offset compensation error Total harmonic distortion 28 between signal terminals each signal terminal to ground 400 kΩ 200 kΩ automatically set by oscilloscope vertical setting 10:1 2:1 ±0.
Version 17.
Version 17.
Version 17.00, November 2016 R&S®RT-ZMA probe tip modules Frequency response Bandwidth Step response Rise time Input impedance Input capacitance with R&S®RT-ZM90 probe amplifier (9 GHz) R&S®RT-ZMA10 tip cable solder-in R&S®RT-ZMA15 tip cable quick connect R&S®RT-ZMA30 browser module R&S®RT-ZMA40 SMA module R&S®RT-ZMA50 extreme temperature kit R&S®RT-ZMA12 tip cable square pin 9 GHz (meas.) 2.5 GHz (meas.) 6 GHz (meas.
Version 17.
Version 17.00, November 2016 R&S®RT-ZA9 probe box to N/USB adapter All parameters are valid for the probe box adapter only. Electrical data Impedance Frequency Return loss Insertion loss Mechanical data Mating cycles Interfaces Connector 1 Connector 2 DC to 12 GHz 12 GHz to 18 GHz DC to 12 GHz 12 GHz to 18 GHz 50 Ω DC to 18 GHz (meas.) ≥ 28 dB (meas.) ≥ 23 dB (meas.) ≤ 0.18 dB (meas.) ≤ 0.22 dB (meas.) RPC-N Rohde & Schwarz probe interface ≥ 500 (meas.) ≥ 5000 (meas.
Version 17.00, November 2016 Ordering information Designation Passive probes 8.0 GHz Transmission Line Probe, 10:1, 500 Ω, 0.3 pF, 20 V (RMS) Incl. signal pin, solder-in (50); ground pin, solder-in (10); signal pin (2); ground pin, pogo (2); signal adapter, square pin (2); ground adapter, square pin (2); SMA(f) to BNC(m) adapter; marker band kit; accessory box; carrying case; operating manual Active probes 1.0 GHz Active Voltage Probe, single-ended, 1 MΩ, 0.9 pF, BNC Incl. spring tip 0.5 mm; solid tip 0.
Version 17.00, November 2016 Designation Differential probes 1.0 GHz Active Voltage Probe, differential, 1 MΩ, 0.6 pF Incl. R&S®ProbeMeter; micro button, R&S®RT-ZA15 external attenuator See R&S®RT-ZD20 for additional equipment included 1.5 GHz Active Voltage Probe, differential, 1 MΩ, 0.6 pF Incl. R&S®ProbeMeter; micro button Incl. signal pin, solder-in (10); signal pin, variable spacing (4); browser adapter; adapter, square pin (2); flex adapter, solder-in 4 cm (1.6 in) and 10 cm (3.
Version 17.00, November 2016 Designation Modular probes 1.5 GHz Modular Probe Amplifier, differential, 400 kΩ, multimode, Incl. R&S®ProbeMeter; micro button Incl. lead 15 cm (5.9 in) (1); Ag plated wire 0.1 mm (0.04 in) (1); Ag plated wire 0.2 mm (0.08 in) (1); solder wire (1); adhesive pads (7); solder-in resistor for R&S®RT-ZMA15 (20); solder lead for R&S®RT-ZMA12/15 (20); marker band kit; carrying case; operating manual 3.0 GHz Modular Probe Amplifier, differential, 400 kΩ, multimode Incl.
Version 17.00, November 2016 Designation Accessories and sets Spare Accessory Set for R&S®RT-ZS10/-ZS10E/-ZS20/-ZS30 active voltage probes Contains: signal pin (10); ground pin, pogo (5); ground pin, solderable, offset (10); ground adapter, square pin (2); marker band kit; mini clip (2); micro clip (2); lead 6 cm (2.4 in) (2); lead 15 cm (5.
Version 17.
Version 17.
Sustainable product design ❙❙ Environmental compatibility and eco-footprint ❙❙ Energy efficiency and low emissions Service that adds value Worldwide Local and personalized ❙ Customized and flexible ❙ Uncompromising quality ❙ Long-term dependability ❙❙ Longevity and optimized total cost of ownership ❙ ❙ Certified Quality Management ISO 9001 Certified Environmental Management ISO 14001 Rohde & Schwarz GmbH & Co. KG www.rohde-schwarz.