User Manual

Putting into Operation
R&S
®
RT-ZS10E/10/20/30
18User Manual 1410.3531.02 ─ 07
4. Adjust the zero position of the waveform using the appropriate function of the
oscilloscope ("AutoZero", "Zero Adjust" or similar).
The waveform is set to 0 V on the horizontal centerline of the oscilloscope.
2.3.2 Micro Button
The micro button provides easy and quick access to important functions of the
Rohde & Schwarz oscilloscope. After a function has been assigned, pressing the
micro button remotely controls this specific function on the base unit. For exam-
ple, "Continuous Run" or "Single" are often assigned to the micro button.
The configuration of the micro button is part of the probe settings of the channel
to which the probe is connected. For more details, see the oscilloscope's user
manual.
2.3.3 Offset Compensation
The offset compensation function can compensate a DC component of the input
signal, even in front of the active amplifier in the probe tip. As a result, the entire
dynamic range of the probe is maintained. This function is useful when measuring
AC signals with high superimposed DC component.
Using the Probe