Key Features and Key Specifications
R&S
®
RTH
6Getting Started 1326.1561.02 ─ 03
Input channels
R&S RTH1002: 2 oscilloscope channels, 1 multimeter
R&S RTH1004: 4 oscilloscope channels
Maximum input voltage
Input CH1, CH2, CH3, CH4 direct (1:1 probe): CAT IV 300 V
Input CH1, CH2, CH3, CH4 via 10:1 probe: CAT IV 600 V, CAT III 1000 V
Meter input: CAT IV 600 V, CAT III 1000 V
Voltage ratings are V RMS (50 to 60 Hz) for AC sine wave and V DC for DC
applications.
Vertical system
Analog bandwidth (–3 dB) of oscilloscope channels:
R&S RTH1002 and R&S RTH1004 with corresponding bandwidth upgrade:
60 MHz, 100 MHz, 200 MHz, 350 MHz, 500 MHz
Rise time (calculated):
R&S RTH1002 and R&S RTH1004 with bandwidth of 60 MHz, 100 MHz, 200
MHz, 350 MHz, 500 Mhz: < 5.8 ns, < 3.5 ns, < 1.75 ns, < 1 ns, < 700 ps
(meas.) respectively
DC gain accuracy:
Offset and position set to zero after self-alignment
Input sensitivity > 5 mV/div: ±1 %
Input sensitivity 5 mV/div: ±1,5 %
Input impedance: 1 M ± 1 % || 12 pF ± 2 pF (meas.)
Input coupling: DC, AC
Input sensitivity: 2 mV/div to 100 V/div
Offset
Offset range depends on input sensitivity:
Input sensitivity ≥ 1 V/div: ±300 V
Input sensitivity 125 mV/div to < 1 V/div: ±(11 V – input sensitivity × 8 div)
Input sensitivity 2 mV/div to < 125 mV/div: ±7 V
Horizontal system
Timebase range: selectable between 1 ns/div and 500 s/div
Modes: normal, roll