Rolls Surrette User Manual

15
SPECIFIC GRAVITY
The specific gravity of electrolyte in a battery is the most accurate measurement
to determine if the cells are in fact, fully charged. The specific gravity should be
constant for 3 hours for an accurate full charge reading. Check the state of charge
as related to specific gravity. Initial charging may take 10 hours. Once the battery
is fully charged, verify the electrolyte level in the cell once more. The fluid should
be 1/4”-1/2” below the vent tube on each cell as shown in Figure 7. Carefully add
distilled water to adjust as needed.
* Check specific gravity regularly to confirm charging parameters are properly
programmed, avoiding possible over/undercharge, resulting in possible cell
damage or battery failure.
Special order batteries may be shipped dry (acid shipped separately). To activate these
batteries, start by removing the vent caps. Using approved battery grade electrolyte
(1.265), slowly fill each cell until the liquid level reached halfway between the plates
and the bottom of the vent well tube. (See Figure 7.) It is important not to over fill
the cells as the acid will expand upon charging. If the cells are too full, the acid will
overflow and require cleanup. Allow electrolyte to saturate into the plates and separators
for at least 90 minutes. The temperature of the electrolyte will rise and the specific
gravity will drop. Once this is complete, place the batteries on charge at the finishing
rate (5%-10% of the 20 hour rate). The rate may be increased if the battery does
not begin to bubble and gas. Do not let the cell temperature exceed 52°C (125°F). If
the temperature becomes excessive or the cells begin to gas vigorously, reduce the rate
of charge. Continue charging until the cell (or cells) reaches within .005 points of the
specific gravity of the filling electrolyte corrected for 25°C (7F). We recommend
allowing the charge to continue for an additional 60 minutes to insure no further rise in
specific gravity. A longer lower charge rate is better than a higher heavier charge.
FIGURE 6:
Temperature
Sensor
TABLE 3:
Specific Gravity
vs State of Charge
% Charge Specic Gravity* (SG)
100 1.255 - 1.275
75 1.215 - 1.235
50 1.180 - 1.200
25 1.155 - 1.165
0 1.110 - 1.130
NOTE: Specific Gravity is dependent on the electrolyte temperature. These values are
for a temperature of 25°C (77°F). To adjust, add/subtract 0.003 for every 5°C (1F)
increase/decrease.
ACTIVATION AND INITIAL CHARGE OF
DRY-SHIPPED FLOODED LEAD ACID BATTERIES