User's Manual

RONDISH STRIP DOOR MONITOR DMS-02
6.5 QUICK SET-UP FOR DOOR STRIPS 001/9 – 001/11)
Strip Door Monitor quick setup procedure
Introducton:
Setting up the DMS-02 strip door monitor has been made easy using the TXP-02(LP) patient tansmitter
for range test.
NOTE:
Do not press the program button on the monitor unit.
Also, please note that maximum range of the unit (rotary switch at zero) is around 5-8 meters.
You should first program the door strip monitors before setting up.
Step 1:
Set rotary switch on monitor/s to mid-point (7). An LED should be "blinking" slowly to indicate "ready".
Step 2:
Fit TXP-02(LP) transmitter unit on wrist and ensure it is switched on using the magnet/Caregiver key (CK-01)
(Please refer to drawing 001/4). If magnetic contacts are fitted and standing near to the door,
you should see another LED flashing to indicate "signal detected".
Step 3:
Move away from the door monitor until this flashing "signal detected" LED stops completely. Next, move back
towards the door until this signal LED flashes again. This point indicates the approximate set up detection
range for the door monitor installation.
Step 4:
If you require more range (longer distance from the door), reduce the setting of the monitor rotary
switch to 6. For less range (shorter distance from door), set the rotary switch to 8.
Step 5:
Continue to adjust the setting of the rotary switch/s and repeat testing until the required range is
obtained.
You may now close the door monitor housing and range setup is complete.
DRAWING No.
001/9V3
TITLE
MODEL No.
Quick Setup Procedure
DMS-02