User's Manual

RONDISH STRIP DOOR MONITOR DMS-02
6.5 QUICK SET-UP FOR DOOR STRIPS 001/9 – 001/11)
Wearing the range test
transmitter wristband(TXP-02(LP)),
walk slowly towards the
door monitor
For setting up detection zone/range, please refer to User Instruction (Page
7 section 3.2)
(a) start with door zone range rotary switch to a mid-way setting (7)
(b) Using the range test transmitter, walk towards the door monitor
(c) When the LEDs flash and a beep is heard, the monitor has detected the
signal from the test transmitter. The distance between transmitter and door
monitor at that time indicates approximate door zone range setting.
note. Only 1 beep (the first) indicates an alarm condition.
higher number = lower range/smaller detection zone.
(d) Using a patient wristband, the door monitor alarm will be triggered. The
electronic coding of the test transmitter indicates range, but does not trigger any
alarm.
Note: Please refer to drawing 001/5. Adjust rotary switch and repeat range test
as necessary until required door zone performance is obtained.
DRAWING No.
001/10V3
TITLE
MODEL No.
Set up the Strip Door Monitor
DMS-02