7.2 Diagnostic messages per NAMUR NE 107
7.2.1 Electronics failure, transmitter
Category
Failure
Cause
An electronics error has occurred.
The device measurement reading is invalid.
Recommended actions
1.
Restart the device.
2. If the condition persists, replace the transmitter electronics.
Related information
Restart the device
7.2.2 Electronics failure, sensor module
Category
Failure
Cause
An electronics error has occurred.
The device measurement reading is invalid.
Recommended actions
1. Restart the device.
2. If the condition persists, replace the transmitter electronics (or the device).
Related information
Restart the device
Service and troubleshooting Reference Manual
September 2022
00809-0100-4418
56 Emerson.com/Rosemount