Users Manual Part 1

SIS configuration error
Cause
It is currently not possible to enable Safety Mode due to other active alerts.
Recommended actions
1.
Clear other active alerts by priority order until this alert is cleared.
2. Change Operational Mode to Control/Monitoring if device is not intended to be
used as safety device.
3. If the condition persists, restore default settings and reconfigure device.
Related information
Rosemount 3408 Safety Manual
Operational mode
Restore to default settings
Configure transmitter using guided setup
Function not supported
Cause
Functionality in the device is enabled, but not supported by this device.
Additional features may be enabled by purchasing an upgrade of the device.
Recommended actions
1. Check that selections for variables (e.g. Primary Variable) are supported by this
device.
2. Turn off functionality not supported by this device.
3. Consider purchasing an upgrade of the device to access additional variables and
functionality.
4. If condition persists, restore default settings and reconfigure device.
Related information
Primary variable
Restore to default settings
Configure transmitter using guided setup
Antenna type configuration error
Cause
The configured Antenna Type is not supported by the device.
Recommended actions
1. Check configuration of Antenna Type.
2. Make sure the configured antenna type matches the physical antenna for the
device.
Service and troubleshooting Reference Manual
September 2022
00809-0100-4418
62 Emerson.com/Rosemount