Users Manual Part 1

Level measurement lost in a manway cover installation
Symptom
The transmitter reports "Level measurement lost" when opening/closing the manway
cover.
Figure 7-12: Symptom
A
B
C
D
A. Level
B.
Time
C. Actual level
D. Level measurement lost
Possible cause
If mounted on a manway cover, the transmitter may lock on a disturbance echo and report
this as surface level when the cover is opened. After closing the cover, the echo is lost and
the transmitter reports "Level measurement lost".
Recommended actions
Before opening the manway cover, disconnect power to the transmitter.
After closing, restart the device.
Move the transmitter to another position.
Related information
Restart the device
Installation considerations
Reference Manual Service and troubleshooting