User's Guide
Table Of Contents
- Contents
- About this manual
- 1 Introduction
- The TankRadar Pro System
- Measurement Principle
- 2 Mechanical Installation
- Socket Requirements
- Free Space Requirements
- Dimensions
- Tools
- Mounting the Rod Antenna
- Mounting the Cone Antenna - PTFE sealing
- Mounting the Cone Antenna - Quartz Sealing
- Mounting the Cone Antenna - Process Seal
- Mounting the Parabolic Antenna
- 3 Electrical Installation
- Identication of Installed Options
- Junction Box
- External Connections
- Connecting to a TRL2 Bus Interface
- Connecting HART devices
- Connecting the Display Panel
- Cables
- Safety
- Power Supply
- Grounding
- System Overview
- Internal Connections
- 4 Pro Setup Software
- System Requirements
- Installing the Pro setup software
- Starting the Pro Setup Program
- Setting a Password
- Logging on to a password protected system.
- Setting measurement units
- Installing a Pro Transmitter
- Viewing Level Data
- Service
- 5 Using the Pro Display Panel
- Operation
- Viewing Level Data
- Installing a Pro Transmitter
- Service
- 6 Tank Distances
- Definitions
- Examples
- 7 Technical Information
- Technical Data
- Display Panel
- Analog Outputs
- 8 Troubleshooting
- Display Panel Error Messages
Saab TankRadar
®
Pro Technical Information
7-1
Edition 2. Ref. No: 306010E
7 Technical Information
Technical Data
Power supply 100-240 VAC or 24-48 VDC
Power consumption Maximum 15 W
Output cabling Twisted and shielded pair, min. 0.5 mm
2
Cable glands EExe approved (CENELEC)
Tank temperature -40-400 °C (-40-750 °F)
Ambient temperature -40-70 °C (-40-158 °F)
Standard measuring range 0-50 m (0-165 ft)
Instrument accuracy ± 5 mm (±0.2”)
Repeatability ±1 mm (±0.04”)
Processors Dual processor design. 32-bit DSP, 50
Mflops
Signal processing FFT, FHAST
Optional disturbance Saab´s Echofixer and
echo handling Multiple Echo Tracking
Level output Display panel, analog 4-20 mA signal with
digital HART,
serial Saab TRL/2 Bus
Ex approval Transmitter Head CENELEC:
(Standard and Gold versions) EEx de [ib] IIC T6.
FM:
Explosion proof Class I, Div. I & 2, Groups
A, B, C and D.