User's Guide
Table Of Contents
- Contents
- About this manual
- 1 Introduction
- The TankRadar Pro System
- Measurement Principle
- 2 Mechanical Installation
- Socket Requirements
- Free Space Requirements
- Dimensions
- Tools
- Mounting the Rod Antenna
- Mounting the Cone Antenna - PTFE sealing
- Mounting the Cone Antenna - Quartz Sealing
- Mounting the Cone Antenna - Process Seal
- Mounting the Parabolic Antenna
- 3 Electrical Installation
- Identication of Installed Options
- Junction Box
- External Connections
- Connecting to a TRL2 Bus Interface
- Connecting HART devices
- Connecting the Display Panel
- Cables
- Safety
- Power Supply
- Grounding
- System Overview
- Internal Connections
- 4 Pro Setup Software
- System Requirements
- Installing the Pro setup software
- Starting the Pro Setup Program
- Setting a Password
- Logging on to a password protected system.
- Setting measurement units
- Installing a Pro Transmitter
- Viewing Level Data
- Service
- 5 Using the Pro Display Panel
- Operation
- Viewing Level Data
- Installing a Pro Transmitter
- Service
- 6 Tank Distances
- Definitions
- Examples
- 7 Technical Information
- Technical Data
- Display Panel
- Analog Outputs
- 8 Troubleshooting
- Display Panel Error Messages
Saab TankRadar
®
Pro Electrical Installation
3-11
Edition 2. Ref. No: 306010E
System Overview
Display Panel
Connect the intrinsically safe Display
Panel to terminals 5, 6 and ground in
the intrinsically safe (EExi) junction box.
Analog Outputs
There are two analog outputs which can
be of passive or active type (external or
internal loop supply). The primary
output has a HART interface.
Connect the primary analog output to
terminals 1 and 2.
Use the EExe junction box for non-
intrinsically safe applications, and the
EExi junction box for intrinsically safe
applications.
Digital Communication
Saab TankRadar Pro can be equipped
with HART interface or Saab TRL/2
Bus interface.
The Saab TRL/2 Bus may only be
connected to the non-intrinsically safe
(EExe) junction box.
Connect the serial data transmission
cables to terminals 1 and 2.
Power Supply
Saab TankRadar Pro operates on 100-
240 V AC or 24-48 V DC. Check the
label on the transmitter housing that the
correct power supply board is installed.
Connect the power supply to terminals
3 and 4 in the non-intrinsically safe
Junction Box (EEx e).
Schematic illustration of Pro trans-
mitter connections
X2
X1
X7
Display Panel
Interface
(DI40)
X7
Level
24500
MENU ITEM GRPH
mm
X3/X5
X3
Junction Box
EEx i
Junction Box
EEx e
HART
®
Interface
1
2
3
4
5
6
5
4
3
2
1
HART modem +
Analog Output
(HM40 + IS40)
Main Board
Power Supply
Unit
(AC40 or DC40)
X4
X4
100-240 VAC
or
24-48 VDC
Extra Analog
Output
(XA40 + IS40)
X3/X6
X6
123
456
78
9
X5