User's Guide

Saab TankRadar
®
ProPro Setup
4-18
Edition 2. Ref. No: 306010E
4 Select the Configuration tab
1 Click the Read button.
Response: the current settings are displayed.
2 In the Tank Geometry box, enter values for Distance Offset (G),
Tank Height (R) and Minimum Level Offset (C).
The Distance Offset (G) is defined as the distance between the
upper reference point and the flange
(
the flange is referred to as the
Tank Radar Pro Reference Point)
. You can use G to specify your
own reference point at the top of the tank. Set G=0 if you want the
flange as upper reference point. G is defined as positive if you use
an upper reference point above the
TR Pro Reference Point.
For
tanks with an ullage plug for hand dipping the
Tank Ullage Refer-
ence Point (dipping mark)
can be used as upper reference point,
but you can use other reference points as well.
The Tank Height (R) is defined as the distance between the upper
reference point (specified by the Distance Offset G) and the lower
reference point (zero level).
The Minimum Level Offset (C) is used to extend the measurement
range down to the tank bottom. C can not be a negative value. C is
defined as the distance between the zero level (
Tank Level Refer-
ence Point
) and the minimum accepted level. If your zero level is
located above the bottom of the tank, set the C distance equal to