Mapping for non-DD based Integration with Host Systems
Appendix C Mapping for non-DD based
Integration with Host Systems
Mapping of alert messages in the HART command 48 additional status . . . . . . . . . . . . page 49
Mapping of device variables index numbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 50
C.1 Mapping of alert messages in the HART command
48 additional status
This outlines the most important alerts in the HART command 48 Additional Status Field for 248
Wireless transmitter. The information in this section can be used by DeltaV
for alert
monitoring, and in the Rosemount 1420 Smart Wireless Gateway for Additional Status mapping
in Modbus, OPC, etc.
A complete list of Additional Status bits is available in the Rosemount 1420 Smart Wireless
Gateway.
Tab le C -1 to Tab le C -3 shows a list of the most important alert messages that may be displayed
in the AMS Wireless Configurator and Field Communication together with the location of the
Alert in the HART command 48 Additional Status field. For recommendation actions refer to
Tab le 6 -1 . on page 35.
To view Active Alerts, from the Home screen, go to Service Tools > Active Alerts.
Table C-1. Failure Alerts (F:)
Message Additional status
(1)
(1) Location of the Alert in the HART command 48 Status field.
Description
Electronics Failure Byte 8 :: Bit 6
An electronics error that could impact the device
measurement reading has occurred
Terminal Block Failure Byte 3 :: Bit 3
A critical failure has occurred in the transmitter's
terminal block
Sensor Failure Byte 3 :: Bit 7
The device has detected an open, short, or too
much resistance for this sensor
Radio Failure
Byte 1 :: Bit 1
Byte 1 :: Bit 7
The wireless radio has detected a failure or stopped
communicating
Supply Voltage Failure Byte 6 :: Bit 3
The supply voltage is too low for the device to
broadcast
Table C-2. Maintenance Alerts (M:)
Message Additional status
(1)
Description
Sensor has Exceeded Limits Byte 3 :: Bit 5
The sensor has exceeded the maximum
measurement range
Terminal Temperature has
Exceeded Limits
Byte 3 :: Bit 1
The terminal temperature has exceeded the
transmitter's maximum range