User's Manual

Table Of Contents
20
Reference Manual
00809-0100-4100, Rev AA
Section 2: Configuration
June 2012
Configuration
2.8.3 Damping
The Damping command introduces a delay in processing which increases the response time of
the transmitter; smoothing variations in output readings caused by rapid input changes. In the
3051 Wireless pressure transmitter, damping only takes effect when the device is placed in high
power refresh mode and during calibration. In normal power mode, the effective damping is 0.
Note that when the device is in high power refresh mode, battery power will be depleted rapidly.
Determine the appropriate damp setting based on the necessary response time, signal stability,
and other requirements of the loop dynamics of your system. The damping value of your device
is user selectable from 0 to 25.6 seconds.
To determine the current damping value:
1. From the Home screen, select 2: Configure.
2. Select 2: Manual Setup.
3. Select 2: Pressure.
4. Select 3: Damping.
2.8.4 Write Protect
The 3051 Wireless has a software write protect security feature.
The view write protect security settings:
1. From the Home screen, select 2: Configure.
2. Select 2: Manual Setup.
3. Select 4: Device Information.
4. Select 6: Write Protect.
Fast Keys
2, 2, 2, 3
Fast Keys
2, 2, 4, 6