User's Manual

Reference
Manual
00809-0100-4045,
Rev
01
Troubleshooting
December
2015
T
r
oubl
esho
o
t
ing
31
1.
L
ocati
on
of
t
h
e
al
e
r
t
in the
HART
command 48 Additional
S
t
atu
s
f
i
e
l
d
.
Table
5-3. Maintenance
Message
Additional
status
(1)
De
s
c
rip
t
i
on
Voltage
c
ond
it
i
on
s
out
of range
B
y
t
e
8::Bit
4
S
upp
ly
v
o
l
t
a
g
e
is
low
and may soon
a
ff
ec
t
device
ope
r
a
t
ion
Pressure
out
of
limi
ts
B
y
t
e
3::Bit
5
Pressure has exceeded
the
m
a
x
i
mum
me
a
s
ure
m
e
n
t
range
Ca
p
a
c
i
t
y
denied
B
y
t
e
12
::
Bi
t
0
Device has
failed
to
acquire
the wire
l
e
ss
communication bandwidth
necessary
to
suppor
t
t
h
e
configured update rates
Sensor
t
e
mpe
r
a
t
u
r
e
beyond sensor
limi
t
s
B
y
t
e
3::Bit
1
Sensor
t
e
mp
e
r
a
t
ure
has exceeded
its safe
ope
r
a
t
ing range
En
vi
ronmen
t
a
l
conditions out o
f
range
B
y
t
e
8::Bit
5
Device is
out
s
id
e
i
t
s
normal
e
n
v
i
ronmen
t
a
l
op
e
r
a
t
ing
c
ondi
t
ion
s
which
may
a
ff
ec
t
accuracy
and/or proper
ope
r
a
t
io
n
1.
L
ocati
on
of
t
h
e
al
e
r
t
in the
HART
command 48 Additional
S
t
atu
s
f
i
e
l
d
.
Table
5-4.
A
dv
i
s
o
r
y
Message
Additional
status
(1)
De
s
c
rip
t
i
on
Da
t
a
b
a
s
e
s
t
o
r
a
g
e
e
rror
B
y
t
e
0::Bit
2
Device has failed
to write to the d
a
t
a
base
me
mor
y
a
t
some
point
in the
past; any
d
a
t
a
written
du
r
i
ng
t
h
i
s
time
may have been
l
o
s
t
Simu
l
a
t
i
on active
B
y
t
e
8::Bit
0
Device is
in
simu
l
a
t
ion mode
and is
not
re
por
t
ing
a
c
t
u
al
in
f
o
rm
a
t
i
on
High power active
B
y
t
e
1::Bit
7
Device is
op
e
r
a
t
i
ng in
a
high power mode; this
i
s
not
recommended
for this device
Non-critical u
s
e
r
da
t
a
warning
B
y
t
e
2::Bit
1
U
s
e
r
-wr
i
tt
e
n p
a
ra
me
t
e
r
does
not m
a
t
c
h its
expe
c
t
e
d value
1.
L
ocati
on
of
t
h
e
al
e
r
t
in the
HART
command 48 Additional
S
t
atu
s
f
i
e
l
d
.