OBM_IB_AC1000f(2)NA10_EN_SW_ver2008.indd 1 Model: AC1000f Blood Pressure Monitor www.rossmax.
Table of Contents: Introduction Introduction .......................................................................................................................................... 3 Cautionary Notes ............................................................................................................................3-5 Notes on Safety................................................................................................................................5, 6 Name/Function of Each Part ....
blood, kidney problems, or for users suffered from stroke, or for unconscious users. 8. This unit is able to detect common arrhythmia (atrial or ventricular premature beats or atrial fibrillation). The ARR, AF(AFib) and PC icons are displayed after the measurement if Atrial Fibrillation and Premature Contraction was detected during the measurement. If ARR, AF(AFib) or PC icons are displayed, you are advised to wait for a while and take another measurement.
• Do not use any cuff other than the models exclusive for this unit. • Do not use this unit on infants. • Do not use this unit on patients using a pump oxygenator. • Do not use an AC adaptor or battery pack no specified for this unit. • Do not use a cellular pone near this unit. • Do not use this unit in a vehicle. • Do not install the parts and/or instruments not specified for this unit. • Do not use a broken power cord or AC adaptor.
Monitor Installation - Trolley (Optional) PARR (Pulse Arrhythmia) Technology Placing Protection Jelly onto the Monitor Pulse Arrhythmia (PARR) technology specifically detects the existence of pulse arrhythmia, including atrial fibrillation (AF, AFib), Atrial and / or Ventricular Premature Contractions (PC), Tachycardia (TACH), and Bradycardia (BRAD). Pulse Arrhythmia may be related to cardiac disorders, needs medical attention and thus early diagnosis is of paramount importance.
Atrial fibrillation most often initially occurs with temporary periods of arrhythmia and may progress to a permanent state of this disorder in the course of time. No matter, whether you intent to safeguard yourself from an undetected AF(AFib) state, or you measure during an ongoing period of active atrial fibrillation, or you measure in between periods of AF(AFib), the PARR technology can be applied at any of these conditions. This unit is able to detect Atrial ) are displayed right fibrillation (AF, AFib).
Pulse Arrhythmia Detection (ARR) Real Fuzzy Measuring Technology Once the occurrence of pulse arrhythmia has been detected in the course of your blood pressure measurement, the icon ARR is displayed. In the case, that the found pulse arrhythmia can be specified by the PARR technology, the ARR icon is accompanied by the specifically detected type of arrhythmia, e.g. PC, AF(AFib), TACH or BRAD.
Display Explanations 1 2 3 4 Loose Cuff Detection 5 6 7 13 8 9 10 11 12 16 14 OBM_IB_AC1000f(2)NA10_EN_SW_ver2008.indd 14-15 14 15 17 1. Arrhythmia Detection (ARR) 2. Premature Contraction Detection (PC) 3. Atrial Fibrillation Detection (AF, AFib) 4. Tachycardia Detection (TACH) 5. Bradycardia Detection (BRAD) 6. Movement Mark 7. Cuff Wrap Detection 8. AC Power Mark 9. Battery Mark 10. Auto Mode 11. Manual Mode 12. Memory Mark 13. Systolic Pressure 14. Diastolic Pressure 15. Pulse Rate 16.
How to Use the POWER Source How to Use the AC Adaptor How to Use the POWER Source 1.Connect the AC adaptor with the AC adaptor jack in the back of the unit. 2.Plug the AC adaptor into the socket. Please use the compatible AC adaptors. (AC adaptors with required voltage and current indicated near the AC adaptor jack.) Warning: • Do not use this unit in places where inflammable gas, such as highly inflammable anesthetic, may be generated or in a high pressure oxygen room or an oxygen tent.
How to Use the POWER Source How to Use the POWER Source Battery life: • You can use the unit for approximately six hundred measurements with one charge. • Approximate life of battery pack is two years. However the battery pack life from each charging may be shortened depending on the state of using. Charging time: • After connecting the AC adaptor, the battery pack will start charging automatically. • While the battery is being charged, the and battery marks turn on and LED shows orange light.
Applying the Cuff Applying the Cuff 1. Select cuff according to arm size: Rossmax Arm Cuff size circumference 3. Place the cuff over the bare upper arm, wrap it with the tube pointing your palm, and the artery mark over your main artery. 4. The edge of the cuff should be at approximately 1.5 to 2.5 cm above the inner side of the elbow joint.
Applying the Cuff MANUAL Mode: with Stethoscope Measurement Procedures AUTO Mode: without Stethoscope Here are a few helpful tips to help you obtain more accurate readings: • Blood pressure changes with every heartbeat and is in constant fluctuation throughout the day. • Blood pressure recording can be affected by the position of the user, his or her physiological condition and other factors.
Measurement Procedures Measurement Procedures Korotkoff sound of Systolic pressure 3. Auto Mode • In Auto mode, mark appears on the display. • Press the START key, the monitor will automatically inflate the cuff slowly to start measurement. • When the measurement is completed, the cuff will exhaust the pressure inside. Systolic pressure, diastolic pressure and pulse will be shown simultaneously on the LCD screen. The measurement is then automatically stored into memory zone.
Setting Deflation Rate Clearing Values from Memory In the Manual mode, select the deflation rate of 2.5 mmHg/sec, 4.5 mmHg/sec, 6.5 mmHg/sec by pressing the key on the back of the unit. Under Auto mode, press and hold the Memory key for approximately 5 seconds, then the data can be erased automatically. How to clean the unit after use Caution: • When cleaning this unit, please unplug the AC adaptor from the electric outlet. You may suffer electric shock.
Specifications Electromagnetic Compatibility Information 1. This device needs to be installed and put into service in accordance with the information provided in the user manual. 2. WARNING: Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the AC1000f, including cables specified by the manufacturer. Otherwise, degradation of the performance of this device could result.
Blood Pressure Log 30 OBM_IB_AC1000f(2)NA10_EN_SW_ver2008.
Warranty Card This instrument is covered by a 2 year guarantee from the date of purchase. The guarantee is valid only on presentation of the warranty card completed or stamped by the seller/dealer confirming date of purchase or the receipt. Batteries, cuff and accessories are not included. Opening or altering the instrument invalidates the guarantee. The guarantee does not cover damage, accidents or non-compliance with the instruction manual. Please contact your local seller/dealer or www. rossmax.com.