Manual

E-M-HF5-V1_25
Rotronic AG
Bassersdorf, Switzerland
Document code
Unit
HygroFlex HF5 Humidity Temperature
Transmitters: User Guide
Instruction Manual
Document Type
Page
29 of 30
Document title
© 2009-2013; Rotronic AG E-M-HF5-V1_25
Conformity with standards
HF52
HF53
HF55
CE / EMC immunity
EMC Directive 2004/108/EG: EN 61000-6-1: 2001, EN 61000-6-2: 2005
EN 61000-6-3: 2005, EN 61000-6-4: 2001 + A11
Solder type
Lead free (RoHS directive)
Fire protection class
Corresponds to UL94-HB
FDA / GAMP directives
compatible
Environmental limits
HF52
HF53
HF55
Storage and transit
-50…+70 °C / -20...+70 °C (models with display), 0…100 %RH, non condensing
Operating limits at electronics
-25 … +60 °C
-10….60 °C
(models with display)
0…100 %RH
non condensing
-40 … +60 °C
-10….60 °C (models with display)
0…100 %RH non condensing
Temperature limits at probe
Depends on probe model
Maximum humidity at sensor
100 %RH up to 80 °C (176 °F)
75 %RH at 100 °C (212 °F)
45 %RH at 125 °C (260 °F)
15 %RH at 150 °C (302 °F)
Maximum air velocity at probe
20 m/s (3,935 ft /min)
Critical environments
Humidity sensor: as per DV04-14.0803.02 - Critical chemicals
10.2 Dew point accuracy
See document E-M-HC2 Probes > Dew point accuracy
11 Accessories
For accessories and parts such as the HW4 configuration software, service cables, calibration accessories
and spare dust filters, please see document E-M-HC2-accessories
12 Supporting documents
Document File Name
Contents
E-M-HC2 Probes-V1
HygroClip 2 (HC2) Humidity Temperature Probes, User Guide
E-M-HC2-accessories
Accessories and parts for probes, indicators and transmitters
E-T-AC3000-DF-V1
AirChip 3000 Description and Main Functions
E-M-HW4v3-DIR
List of the HW4 manuals
E-M-HW4v3-Main
HW4 software version 3: General instructions and functions common to all
devices
E-M-HW4v3-F2-005
HW4 software version 3: Device Manager - HF5 transmitter