Data Loggers for Humidity and Temperature (only for V2.0)
SHORT INSTRUCTION MANUAL
General Description
Congratulations on your new ROTRONIC HL-1D (humidity and temperature) / TL-1D (tem-
perature) data logger. Please read these short instructions carefully before using the device.
These short instructions are limited to a description of the main functions and installation
of the device.
The HL-1D data logger can log up to 32,000 pairs of measured values (relative humidity and
temperature) in a range from 0 to 100 %RH and -20 to 70 °C.
The TL-1D data logger can log up to 64,000 temperature measured values in a range from
-20 to 70 °C.
Dimensions / Connections
Battery
compartment
(BAT.)
USB-Mini port
(settings and data download)
Inserting / Replacing the Battery
ROTRONIC HL-1D/TL-1D data loggers are powered by a CR2 battery.
To insert/replace the battery, open the BAT lid with a coin and insert the battery with the right
polarity (+/-). When closing the lid, make sure that the O-ring seal is positioned correctly.
Mechanical Installation / General Recommendations
Relative humidity is very temperature-dependent. The installation site can therefore have a
signifi cant infl uence on the performance of the device.
Follow the guidelines below to ensure optimum performance:
a) Select a representative installation site: install the device at a point where the humidity,
temperature and pressure conditions are representative of the environment that is to be
measured.
b) Ensure there is suffi cient air movement at the device: an air velocity of at least 1 meter
per second accelerates and facilitates adaptation of the probe to changing temperatures.
c) Avoid: (1) placing the device too close to heating elements, cooling coils, cold or hot walls,
in direct sunlight, etc. (2) placing the probe too close to steam injectors, humidifi ers,
in direct precipitation. (3) unstable pressure conditions with high air turbulence. (4)
accumulation of condensation at the contact wires of the sensor.
Mounting the Data Logger
The device can be placed at any point or mounted on a wall. Wall mounting: pull out the
two wall holders at the back of the device. Drill two holes and screw in screws. Then hang
the device on the screws.
COMM: Appears when the device is connected to a computer
MIN/MAX/AVG: Press button briefl y to show MIN/MAX/AVG of the stored
measured values (only works in LOG mode)
FULL: Appears when the memory is full
LOG: Appears when the logging function is active
: Appears when the battery is low
°C/°F, %RH: Temperature and humidity unit
Hi, Lo: Appears when the temperature or humidity is above the High Alarm
setting or below the Low Alarm setting
On/Off / Data Logging
Press button briefl y to switch the device on/off
(the device cannot be switched off when in logging mode).
To start or stop data logging, press the START/STOP button for a few seconds.
The right LED fl ashes green every ve seconds during data logging.
The left LED ashes red every ve seconds when values exceed/drop below the alarm settings
or when the battery is low.
Display
Settings / Viewing Recorded Data
A PC running the ROTRONIC software HW4 is needed to make settings and download/view
the recorded data. The device and PC are connected to each other with a normal USB to
USB-Mini cable (AC003). The cover of the USB port on the device can be opened with a coin.
More information regarding the setting of the HW4 software can be found in the help-fi le
of the device manager.
CAUTION: Time must be synchronized when fi rst using the logger.
Troubleshooting
Error Problem Solution
E02 Measured temperature value
is too low
Place the device in a normal environment
E03 Measured temperature value
is too high
Place the device in a normal environment
E31 Temperature sensor damaged Send the device in for repair
E04 Temperature measurement error Place the device in a normal environment
E11 RH calibration error Repeat humidity calibration
E33 Circuit damaged Send the device in for repair
Technical Data
Type of battery 1 x CR2
Battery life 3 years (at logging interval of 1 hour)
Battery charge indicator Yes (HW4 software and display / LED indicator)
Range of measurement /
application
0…100 %RH / -20...70 °C
Accuracy at 23 °C ±5 °C ±3.0 %RH / ±0.3 °C
Storage capacity HL-1D: 32,000 data points (%RH + °C)
TL-1D: 64,000 data points (°C)
Logging interval 30 s to 24 h
Display LCD, 2 lines, resolution of 1 decimal
LED indicator 2 LEDs
Right LED ashes green during data logging
Left LED fl ashes red when limits broken
or low battery
Housing protection grade HL-1D: IP67 (electronics)
TL-1D: IP67
Weight 85 g
FDA/GAMP compatibility Conformity
Dimensions 90 x 60 x 23 mm
Software HW4 V3.4 or higher
Included HL-1D / TL-1D
CR2 3V lithium battery
Short instruction manual
Calibration certifi cate
Log-Interval
Storage capacity reached after
TL-1D HL-1D
30 sec 22 days 5 hours 11 days 2 hours
5 min 222 days 111 days
30 min >3 years 1 year 301 days
45 min >5 years 2 years 270 days
60 min >6 years >3 years