User guide

E-M-HP22-V2_13
Rotronic AG
Bassersdorf, Switzerland
Document code Unit
HygroPalm HP22-A hand-held indicator:
User Guide
Instruction Manual
Document Type
Page
10 of 16
Document title
© 2010-2012; Rotronic AG E-M-HP22-V2_13
4.9 Internal menu
Note: Unauthorized access to the menu can be prevented by disabling the “display menu” setting
(use the HW4 software > Device Manager > Display)
Main Menu
Menu Items
Selections / Information
Notes
Device Settings
Units
Metric / English
Back Light
Key Press / On / Off
Display backlight mode
Contrast
LC display contrast adjustment
Trend
On / Off
Trend indication on the display
AutoPwrOff
OFF
1 min
5 min
15 min
30 min
45 min
60 min
The HP22 powers itself down after
a user specified time period
Device Information
Version
Firmware version
Serial Nbr
Serial number
Address
RS-485 address
Type
Device type
Name
Device name
User defined
Probe Information
Version
Firmware version
Serial Nbr
Serial number
Address
RS-485 address
Name
Device name
User defined
SensorTest
Humidity sensor status
Off / Good / SQ-Tuned / Bad
Record
On / Off
Data recording by the probe
(max. 2000 values)
Humidity Adjust
RefValue
Humidity reference value
± 0.1 %RH steps
Acquired
Number of cal. points in probe
memory
<Acquire>
Save cal. point to probe memory
<Delete>
Erases all calibration points
<Adjust>
Effect depends on number of
calibration points
Temperature Adjust
RefValue
Temperature reference value
± 0.1 ˚C steps
<Adjust>
1-point adjustment only (offset)
o Record: both the recording mode (start / stop and the log interval cannot be changed from the menu and
are as configured with the ROTRONIC HW4 software