Page 23 of 75
Document title
Timing Tolerances (items 10 & 11)
For each program, temperature and humidity stabilisation values can be defined (items 10 & 11). Once
the chamber conditions are within this tolerance, the step duration timing begins. If set to zero, no
stabilisation criteria are applied and the timing of the step duration begins immediately.
Temperature only Program (item 12)
Humidity timing and control can be removed from a program using the ‘Temperature Only’ button.
Programmer Screen
Click the ‘Prog’ button (item 13) to return to the Programmer screen.
3.4.5 Settings Screen
1) Water level display
2) Desiccant condition display
3) Settings menu
4) Updates menu
5) Support menu
The water level and desiccant level indicators show approximate levels only. The desiccant level
indicator is updated only when the HygroGen2 is drying at more than 15% power and will take a few
minutes to update. This is to ensure the desiccant sensor is reading the correct value. See section 8.2
for more details.
The drop-down menus (items 3, 4 & 5) allow navigation to screens that are detailed later in this
document.
Figure 12: Settings Screen