Owner's manual

E-M-HG2-S-V2.1
Rotronic AG
Bassersdorf, Switzerland
Document code Unit
HygroGen2: Humidity and Temperature Generator
with AutoCal/AutoCal+, Remote Control and
Range Extensions options.
Instruction Manual
Instruction Manual for Software
Version 2.1
Document Type
Page 38 of 75
Document title
Extremes of Temperature and Humidity
As with any other reference or instrument under test that is inserted through the door, stem conduction
should be considered at the extremes of the HygroGen2’s measurement range.
A recommendation for conditions that are higher than 50°C, 90%rh, or 60°C, 80%rh is that the entire
RP2 measuring head is mounted on the inside of the door. At these extremes, condensation will form at
cold spots that can be introduced if there is thermal conduction with the ambient conditions.
Figure 28: Probe mounted on inside door surface
An HG2-B8-9 bung which has a 9-17mm cable gland can be used to secure the head cable from the
outside of the door and provides further insulation.
© 2014; Rotronic AG E-M-HG2-S-V2.1