Owner's manual

E-M-HG2-S-V2.1
Rotronic AG
Bassersdorf, Switzerland
Document code Unit
HygroGen2 Humidity and Temperature Generator:
Instruction Manual
Appendix
Instruction Manual for Software
Version 2.1
Document Type
Page 60 of 75
Document title
B1: Validation requirements and accessories
For an ISO 17025 calibration system, you will need to show evidence of qualifying your own system,
and permanently logging temperature gradients within the chamber.
For this purpose, ROTRONIC has developed a special door with detachable mapping rig:
And a PRT mounting ring that can be fitted permanently into the chamber for continuous monitoring:
We recommend the MBW T12 multi-channel thermometer for performing these tasks, which when
used with the MBW/RHS dew point mirrors provide a calculated RH map of the entire space.
http://www.mbw.ch/product_T12.php
Figure 40: HG2-PRTring
Figure 39: HG2-D888888-Map
© 2014; Rotronic AG E-M-HG2-S-V2.1