Use cable AC3006 to connect the service connector of the HygroMet MP to a USB port of a PC with
the ROTRONIC HW4 software installed. Note that the ROTRONIC USB driver must be installed on
the PC as explained in the HW4 manual E-M-HW4v3-Main
Power the HygroMet MP from a DC voltage source. Be sure to power the HygroMet MP during the
entire firmware update process.
Copy the firmware update file from the ROTRONIC website to the PC.
Start HW4 software on the PC and search for the HygroMet MP (HW4 Main Menu Bar > Devices and
Groups > Search for USB Masters).
After finding the HygroMet MP, expand the device tree to see the HygroMet MP functions. Select
Device Manager. In the Device Manager menu bar select Tools > Firmware Update. For instructions
see document E-M-HW4v3-F2-014