MEASUREMENT LOOP VALIDATION
Functions
Description
Simulator mode
Used to make the HygroMet generate fixed values for the
humidity, temperature and calculated parameter. Can be
configured, enabled or disabled
DEVICE SAFEGUARDS
Functions
Description
Device write protection
Used to protect the HygroMet with a password to prevent
unauthorized digital access by a digital user. Can be configured,
enabled or disabled
PROCESS PROTECTION AND PROTECTION OF OTHER DEVICES
AirChip 3000 Functions
Description
Limit humidity output to 100 %RH
This probe function is used to prevent the humidity signal from
exceeding 100 %RH when condensation forms on the sensor.
Can be enabled or disabled
Out-of-limit value alarm
Used to specify the normal range for humidity, temperature and
the calculated parameter depending on the user application.
Can be configured, enabled or disabled
Out-of-limit values trigger a digital alarm which can be also be
seen on the optional display
Bad sensor alarm
This is a built-in probe function. Cannot be disabled
A bad humidity or temperature sensor triggers a digital alarm,
provided that the HygroMet is set to monitor probe alarms
Fail safe mode
Used to specify a "safe" fixed value for humidity and for
temperature (HygroMet or probe) in the event of:
o Loss of communication with the probe (HygroMet function)
o Sensor failure (probe function)
Can be configured, enabled or disabled
4.3 Factory default settings
Configurable Settings
Applicability
Factory default
Unit system (Metric or English)
2-wire, 3-wire
As per ordering code
Analog signal type (4…20 mA or other)
3-wire
As per ordering code
Psychrometric calculation
2-wire, 3-wire
As per ordering code
Fixed pressure value
2-wire, 3-wire
1013.25 hPa or 29.92 In Hg
Output 1 parameter, scale and unit
2-wire, 3-wire
As per ordering code (%RH or other)
Output 2 parameter, scale and unit
2-wire, 3-wire
Temperature, unit as per ordering code
Communication protocol
3-wire
RO-ASCII
RS-485 address
3-wire
0
Device name
2-wire, 3-wire
Instrument model