Use Instructions

Table Of Contents
Input water connection
1x John Guest PM Acetal fitting 8 mm
Produced water quality
Clinical laboratory reagent water, conductivity ≥ 1MΩ/cm
Installation requirement
Installation by Merck-Millipore Field Installation Engineer is
required
Waste
RPI Waste connection
1x John Guest PM Acetal fitting 10mm (OD)
AFS-24 Waste connection
1x John Guest PM Acetal fitting 6 mm (OD)
Concentrate cubitainers
Cubitainer positions
2
Cubitainer orientation
Upright
Cubitainer volume
10L
Maximum Cubitainer
dimensions
H 245 x W 245 x D 245 mm
Cubitainer detection &
identification
NFC
Location RFID at cubitainer
Rear
Concentrate pick-up tube
To be determined with customer
Monitoring process:
Monitoring method
Control cycle using RPI-Check
RPI-Check temperature
Acclimated to environment temperature (>24 hours)
Frequency
≥1x per 24 hrs
Control process start
Manual start and time scheduled
Internal reagent storage:
Storage volume
10 liters
Storage temperature
Displayed in GUI and monitored (not controlled)
Performance characteristics and specifications
16