Reagent delivery port details:
Number of output ports
4
Output port outer diameter
John Guest PM Acetal 6 mm (OD)
End point filtration
0.22 µm
Expected priming vacuum
~ 33 kPa (250 mmHg) by analyzer
User interface
Human machine interface
Full color touch screen
Display specifications
154x90mm (7"), 1024x600 resolution, 85˚ viewing angle
User access levels
Operator / Lab manager / Lab technician / Field Service Engineer
(FSE)
Languages
English, Dutch, French, Italian, German, Spanish
Intended use:
Use environment
In-Vitro Diagnostic laboratories
Intended operators
Trained IVD lab personnel
Cubitainer switching
Intra lot cubitainer switching
Uninterrupted production
Lot-to-lot cubitainer switching
When storage is empty, prime with new lot.
Analyzers background cycles recommended.
Operational
Maximum ready to use
reagent storage time
To be determined by customer
Mobility
4 wheels
Sanitation
Reconstitution rinse cycle
Timer initiated, once every 24 hrs.
Rinse medium
Clinical laboratory reagent water, resistivity≥ 1MΩ/cm
Performance characteristics and specifications
MRN-184-EN Version 6 (05-01-2022)