Remove side panels
1. Undo the latches from a side panel on the inside of the RPI.
2. Now gently push the side panel outwards.
3. Lift the side cover up out of the lower supports with use of the grip.
Place the side cover in a suitable location until it is placed back.
4. Repeat Steps 1 to 4 for the other side panel.
Reverse the steps to put the panels back.
11.3 - Replacing the Vent filter
Only to be performed by a trained Lab Technician or FSE. Be aware of liquid spilling, use a cloth
to prevent liquid dripping in the instrument.
The vent filter (SQWLV040002) must be replaced every 3 months. This is indicated before the due date and
the message will be shown until replacement has been done.
Maintenance
MRN-184-EN Version 6 (05-01-2022)
75