F59 RPI-Check bottle open life expired Open life expiration is detected Replace RPI-Check bottle
F60 When starting control, open life expired After start of control step shelf
life expiration is detected
Replace RPI-Check bottle
F61 Shelf life one concentrate cubitainer
expired
Shelf life expiration is detected Replace cubitainer
F62 Shelf life both concentrate cubitainers
expired
Shelf life expiration is detected Replace cubitainers
F63 RPI-Check bottle shelf life expired Shelf life expiration is detected Replace RPI-Check bottle
F64 When starting control, shelf life expired Shelf life expiration is detected Replace RPI-Check bottle
F65 Door opened without logging in. Please
close the door, or log in by pressing confirm
Unauthorized action Close door/drawer or log in
F66 Internal communication error Menu not found Navigate to other menu
Retry action that caused the error
F67 Warning: the level of ready to use
reagent is too low
Low level of production. Risk of
air bubbles. Will occur after lot
switch.
In case of recurrence:
Replace filters of WPU
Call for Service
F68 Warning: cubitainer tag could not be
read
Damaged NFC tag or NFC
reader failure
Enter barcode manually
F69 Warning: The instrument is not
configured for this type of cubitainer
Not supported type of
cubitainer placed
Replace cubitainer with correct
producer ID
F70 Warning: Entered barcode information is
not valid
Mistyping or incomplete enter
barcode
Enter correct barcode
F71 Database failure System error Call for service