SiteStand® Mobile Docking Station Service Manual
Manufactured for SonoSite, Inc. 21919 30th Drive SE Bothell, WA 98021-3904 USA Telephone: 1-888-482-9449 or +1-425-951-1200 Fax: +1-425-951-1201 SonoSite European Headquarters Baystrait House, Station Road Biggleswade SG18 8AL UK Telephone: +44-1767-313-117 Fax: +44-1767-312-400 CAUTION: ` United States federal law restricts this device to sale by or on the order of a physician. “SiteCharge,” “SitePack,” “SiteStand,” and “SonoHeart,” are trademarks of SonoSite, Inc.
iv
Table of Contents 1 Introduction 1 1.1 SiteStand mobile docking station..............................................1 1.2 SiteStand System Display .........................................................1 1.3 Setup and Operation..................................................................2 1.4 Specifications............................................................................3 1.4.1 SiteStand mobile docking station .......................................3 1.4.
A Schematic and Parts List 15 A.1 Schematic and Wiring Diagram..............................................15 A.2 Replacement Parts List ...........................................................15 A.3 Ordering Replacement Parts ...................................................15 A.4 Cabling Diagram.....................................................................16 A.5 Schematic Diagram, Docking Stand PCB ..............................20 A.6 SiteStand Station Assembly.............................
CHAPTER 1 Introduction Before servicing the SiteStand® mobile docking station (SiteStand), read and be familiar with the information in this manual. 1.1 SiteStand mobile docking station The SiteStand station (Figure 1.
1.3 Setup and Operation For setup and operation of the SiteStand mobile docking station, refer to the SiteStand Mobile Docking Station User Guide (P01542-06). 15 Inch Display SonoSite system Sleeve SiteStand mobile docking station Printer bracket and printer Basket VCR bracket and VCR Height adjustment pedal Locking levers Figure 1.
1.4 Specifications 1.4.1 SiteStand mobile docking station The SiteStand mobile docking station operates on either 50 or 60-Hertz alternating current (AC) power (Table 1.1). Table 1.1 SiteStand station AC Electrical Power Specifications INPUT VOLTAGE NOMINAL CURRENTS (AMPS) OUTPUT VOLTAGE NOMINAL CURRENTS (AMPS) AC 100 -240 1.0 - 0.50 + 16 Vdc AC 100 -240 50/60 Hz 2.8 2.0 (2x) 1.0 - 0.50 + 16 Vdc 2.8 DEVICE SiteStand 50/60 Hz AC power adapter AC 100 -240 50/60 Hz 1.4.
4 SiteStand Mobile Docking Station Service Manual
CHAPTER 2 Safety Please read this information before using this ultrasound system. It applies to the ultrasound system, transducers, peripherals, and accessories. A WARNING describes precautions necessary to prevent injury or loss of life. A CAUTION describes precautions necessary to protect the products. 2.1 Electrical Safety WARNINGS: ` To avoid the risk of electrical shock, use commercial grade peripherals recommended by SonoSite on battery power only.
CAUTIONS: ` Electrostatic discharge (ESD), or static shock, is a naturally occurring phenomenon. ESD is common in conditions of low humidity, which can be caused by heating or air conditioning. Static shock is a discharge of the electrical energy from a charged body to a lesser or non-charged body. The degree of discharge can be significant enough to cause damage to a transducer or an ultrasound system.
2.4 Biological Safety Observe the following precautions related to biological safety. WARNING: ` To avoid injury to user, patient, or bystander, lower the stand height prior to transporting and pull the stand over thresholds. Failure to lower the stand before pulling over thresholds when the 15 inch display is attached could result in the stand becoming unstable, tipping, or falling over. Symbol The following symbol is found on the 15 inch display.
8 SiteStand Mobile Docking Station Service Manual
CHAPTER 3 3.1 Servicing the SiteStand Introduction Before servicing the SiteStand mobile docking station, read and be familiar with the information in this manual. The SiteStand assembly drawings and part lists are located in Appendix A, Schematic and Parts List on page 15. 3.1.1 Contacting SonoSite Technical Support For technical support or to order spare parts, do one of the following: • • • • • 3.2 For U.S.& Canadian customers, call 1-877-657-8118. For all Worldwide customers, call +425-951-1330.
Table 3.1 SiteStand Station Troubleshooting OBSERVATION The system battery doesn’t charge in the SiteStand station. POSSIBLE CAUSE If the charge function of the system is acceptable, the possible causes are: • the power supply has failed or become disconnected • the dock PCBA located in the sleeve assembly has failed • the docking connectors on the sleeve assembly are broken or bent.
3.3 Disassembling the SiteStand mobile docking station Disassembly of the SiteStand station is dependent on the repairs required. The following order of disassembly is recommended in order to prevent damage to the SiteStand station and facilitate efficient repairs. The service provider uses professional judgement in altering this order to provide repairs to various parts of the SiteStand station. 3.3.1 Required Materials and Equipment • • • • 3.3.
3.3.4 Repairing or Replacing the Front Panel Connectors and the Power Supply 1. To access the connectors and the power supply use a #1 philips screwdriver to remove the six connector cover philips screws. Pull the connector cover off and set it aside. 2. Extend the gas spring to its maximum height by pressing the foot pedal. 3. Remove the blade cap by gently prying it up from the sides with a flat blade prying tool. There are no screws holding it in place.
3.3.5 Replacing the SiteStand station Fuse RS-232C communication connector Video output connector Audio output connector Power output connectors Printer and VCR control connector Video output connector AC power input connector Fuse drawer To replace the electrical fuse: 1. Disconnect the AC line cord from the hospital-grade electrical outlet. 2. Disconnect the AC line cord from the bottom of the SiteStand mobile docking station. 3.
3.4 SiteStand Mobile Docking Station Performance Tests 3.4.1 Setup 1. Install any SonoSite ultrasound system (without a battery) into the SiteStand mobile docking station. 2. Apply AC power, 120V 60Hz to SiteStand station “AC in” connector. 3.4.2 AC Power Verification Use an AC Tester to verify both IEC output connectors are wired correctly. 3.4.3 Docking Video 1. Turn on the system. 2. Connect a Test Monitor to each of the video outputs on the docking stand. 3.
APPENDIX A Schematic and Parts List This section contains a list of field-replaceable parts. A.1 Schematic and Wiring Diagram This appendix provides a schematic diagram, a wiring diagram, and a cable diagram for the SiteStand station and the docking stand PCB. A.2 Replacement Parts List This appendix also provides tables that contain all the replaceable parts for the SiteStand station. All quantities are one unless otherwise noted. A.
A.4 Cabling Diagram B/W Printer Power Output - IEC IEC Input AC Receptacle (male) Receptacle (female) Auxiliary Power Output - IEC Receptacle (female) Power Supply Input Fuses: Hot & Neutral 6.
Docking Station Fingers Table A-1 shows the 14 interface docking fingers that engage when the UI unit is put in place in the docking station. These signals are then routed to sub circuits or pass through the Docking Station PCB to the mixed signal coax connector described in table A-1. From the connector the signals are routed through a cable to their appropriate interfaces in the base of the Docking Station. Table A.
External Video Connector The video supplied to the External Video Connectors on the Docking station is identical to the video generated from the UI Module. This video is routed out to the appropriate output jack, refer to Table 6-1 for signal conditions Table A.2 SIGNAL External Video PIN(S) I/O LEVEL Ext_Video O 1 Vpp GND - 0V DESCRIPTION Composite Video Output Signal Return External RS-232 Com Connector The connection to the 9 pin D-Sub connector for the RS-232 is shown in Table A.3. Table A.
Video Printer Control This is a stereo jack that provides print control and status (Busy) for the video printer Refer to Table A.4 for details. Table A.4 Video Printer/ VCR Control SIGNAL PIN(S) I/O LEVEL DESCRIPTION Control_line Tip(2) O +5V Printer Control Stat_Frame Center Ring(3) O +5V Printer Status GND Outer Ring(1) - 0V Printer Return AC Adapter Interface The AC Adapter Interface consists of a module mounted on the bottom of the docking station.
A.
A.
TThe parts list below is for the SiteStand Station Assembly shown on page 21. Table A.7 22 SiteStand station Assembly Parts List FIND NUMBER ITEM NUMBER 1 P01607 assembly, blade 2 P01619 subplate, body, top 3 P00589 blade cap 4 P02547 assembly, sitestand bearing and housing 6 P01622 tube, lift 7 P01623 bushing, clocking 8 P01624 base, machined 9 P00994 pedal, actuator 10 P00693 caster 11 P00694 caster, nonlocking 12 P01501 screw, shcs,10-32 x .
Table A.7 SiteStand station Assembly Parts List, Continued FIND NUMBER ITEM NUMBER 30 P00590 cover, connector, with audio 31 P01566 screw, phpls, 6-32x.50lg, zinc-chromate 32 P00976 bumper, elastomeric 33 P00908 sitestand support bracket 34 P01788 rivet, 3/16 x .50 open end, stl 36 P01500 nut, sleeve, non-locking 37 P01625 screw, shcs,10-32 x .25 lg, alloy 38 P01788 washer, lock, internal tooth, 3/8, zinc 39 P02022 screw, flat head, 10-32 x .
A.7 SiteStand Station Sleeve Assembly Table A.
APPENDIX B Service Event Report The Service Event Report provides information about product failures to the manufacturer and to authorized service facilities, which provide approved warranty services for SonoSite products. For all repairs completed, complete the form and return a copy of it to the following address: SonoSite, Inc. Technical Support 21919 30th Drive SE Bothell, Washington 98021-3904 telephone: 1-877-657-8118 (U.S.
Service Event Report Service Provider Name: Date: Company: Address: Phone Number: Fax Number: E-mail address: Device Description Name: Serial Number: Part Number: Lot Number: Software Version: Other Identifiers: Revision: Event Description Diagnosis Service Performed Performed By: Actions: Date: Parts Removed Part Name Part Number Serial Number Lot Number Rev Replaced By Parts Installed Part Name Part Number Serial Number Lot Number Rev Replaced By Tests Performed (attach test data)