Operation Manual
Sagab by Elma
911
25
SYMBOLS AND ANNUNCIATORS
Continuity
BAT Low Battery
Diode
DATA HOLD Data Hold
AUTO AutoRanging
AC Alternating Current or Voltage
DC Direct Current or Voltage
Specifications
The instrument complies with: EN61010-1.
Insulation: Class2, Double insulation.
Overvoltage category: CATIII 600V, CATII 1000V.
Display: 2000 counts LCD display with function indication.
Polarity: Automatic, (-) negative polarity indication.
Overrange: “OL” mark indication.
Low battery indication: The “BAT” is displayed when the battery voltage drops below the
operating level.
Measurement rate: 2 times per second, nominal.
Auto power off: Meter automatically shuts down after approx. 15 minutes of inactivity.
Operating environment: 0
o
C to 50
o
C (32
o
F to 122
o
F) at < 70 % relative humidity.
Storage temperature: -20
o
C to 60
o
C (-4
o
F to 140
o
F) at < 80 % relative humidity.
For inside use, max height: 2000m
Pollution degree: 2
Power: One 9V battery , NEDA 1604, IEC 6F22.
Dimensions: 150 (H) x 70 (W) x 48 (D) mm
Weight: Approx.: 255g.
Accuracy is given at 18
o
C to 28
o
C (65
o
F to 83
o
F), less than 70 % RH
DC Voltage (Auto-ranging)
Range
Resolution
Accuracy
200.0mV 0.1mV +0.5% of rdg + 2 dgts
2.000V 1mV
+1.2% of rdg + 2 dgts
20.00V 10mV
200.0V 100mV
600V(CATIII) 1V
+1.5% of rdg + 2 dgts
1000V(CATII) +1.5% of rdg + 4 dgts
Input Impedance: 10MΩ.
Maximum Input: 600V dc/ac rms.(CATIII)
Maximum Input: 1000V dc/ac rms.(CATII)