Use Instructions

SAIT‐IFU‐GEMS‐HRev.1.1(2019‐11) Page45of79
6.7.2.Assistcontrolarea
6.7.2.1.Setthestandpose
Makesurethathipangleisasclosetoneutralaspossiblewiththepatient standing naturally.Ifthepatientis
walkingwiththeirbackbent,theirneutralposturewillalsobestandingwithbackbent.
Incorrectsettingofthestandposemaycauseuncomfortableassistanceordegradeclinicaleffectiveness.
Makesurethatboththeleftandrightjointangleoptionsareselectedwithinthemonitoringoptiondialog.
Resetthehipanglesensorwhilethepatientisstandinginaneutralposturewithhipsstraightifpossible,orwith
comfortablepostureotherwise.Thiswillbeusedasthereferencepositionforcalculatingflexionandextension.
Pressthe“SetStandPose”buttontoresetthehipangletozeroandtocalibratethereferenceposture.
Ifthestandposeisnotset,the“START”buttonwillnotbeactivated.
Makesurethatbothlefttherightjointanglesareselectedinthemonitoringoptiondialogtoseebothjoint
anglesinthegraph.
