Use Instructions

SAIT‐IFU‐GEMS‐HRev.1.1(2019‐11) Page53of79
6.8.Safetyfeatures
6.8.1.MitigationofrisksrelatedtoafullydischargedLithiumIonbattery
TheLiionbatteryusedinthisdevicehasariskofexplosionandofmaterialleakagewhencompletelydischarged.To
prevent this, the power control module (PCM) circuit is designed to fundamentally cut off the power before the
batterydischargesfully.Inaddition,thebatteryleveliscontinuouslymeasuredanddisplayedontheapplication.
Statusofthedeviceandrecommendeduseraction
AnalertwillappearontheassociatedtabletPCifthebatterylevelisbelow20%and10%,andthebatterylevelwill
beshowninred.
Whenthebatterylevelisbelow20%,thephysicaltherapistisrecommendedtochargethebatterywithin5minutes.
Whenthebatterylevelisbelow10%,itisrecommendedthatdeviceoperationbestoppedimmediately,andthat
thedevicebechargedbeforefurtheruse.
Asafinalwarning,ifthebatterylevelisalmostzero,thenwithout further warning messages, the low battery
warningsoundwillbeplayedfromthedevicespeaker,anddeviceassistancewillbestoppedimmediately.
Errormessageexamples