6.8.2.Mitigationofrisksduetoexcessivepoweroutput
Inordertoavoidanexcessivetorquebeingappliedtothepatient,thedeviceisdesignedtonottotransmitatorque
of more than12 Nm ±15% to thepatient. When the assistive forcereaches thepredeterminedmaximumforce,
thenthedevicewillceaseapplyingtheassistiveforce.
Statusofthedeviceandrecommendeduseraction:
Ifassistiveforceisceasedduetoanovertorquesituation,theassociatedtabletPCwilldisplayanerrormessageof
EXCEEDEDTORQUELIMIT.”
Whenthismessageisdisplayedandthedeviceisstopped,thephysicaltherapistmaytrytostartthedeviceagainby
pressingthe“START”onthetabletPC.Iftheproblempersists,rebootthedeviceandthetabletPC.Ifthereisstillthe
problemwiththedevice,turnoffthedeviceimmediatelyandcontactthemanufacturer.
Errormessageexamples
