Use Instructions

SAIT‐IFU‐GEMS‐HRev.1.1(2019‐11) Page56of79
6.8.3.Mitigationofrisksduetoexcessivejointanglemovement
TheGEMS‐Hjointshavemechanicalstopperstorestrictover‐extensionofjoints,aswellassoftwarelimitsonjoint
movement.Themechanicalstoppersaredesignedtoensurethatthehipjointsstaywithinthespecifiedrangeeven
inthecaseofamalfunctioningdevice.Themechanicalstoppersaresettoamaximumof110degreesforflexionand
amaximumof45degreesforextension.Thedelayedoutputcontroller(software)islimitedtoamaximumof100
degreesforflexionandamaximumof40degreesforextensioninnormaluse.
Statusofthedeviceandrecommendeduseraction:
Whenthejointanglereachesthejointanglelimit,thedevicealertsthephysicaltherapistbydisplayingawarning
message of EXCEED JOINT LIMIT”. When the joint angle reaches these joint limits, the device ceases applying
assistive force. Thereafter,the physical therapist mayreset the device byaskingthepatientto stand with upright
postureandtappingthe“SetStandPose”buttononthetabletPC.Pressingthe“START”buttonwillthencausethe
devicetostartprovidingassistiveforceagain.
Errormessageexamples
