6.8.5.Mitigationofrisksduetoabnormalactuatormodulestatus
If the actuator module does not work properly during use, the controller may generate abnormal power. The
actuator module is monitored continuously during treatment, andwhenanabnormalityisdetected,theassistive
forceisturnedoff.
Statusofthedeviceandrecommendeduseraction:
Shouldabnormalactuatorstatusbedetected,thetabletPCwilldisplaytheerror“DETECTEDABNORMALACTUATOR”
Theactuatorstatusisconsideredabnormalwhendatatransmittedfromtheactuatordoesnotmatchtheexpected
signals,forexample,whenthedevicedetectspatientmovement,butactuatorreadingsareconstant.Thismayresult
fromdisconnectedcommunicationlines,actuatorfaults,and/orsoftware failure. The physical therapist should
attempttofixtheproblembyrebootingboththedeviceandthetabletPC.Ifthiscorrectstheerror,thetabletPCwill
display“NORMALOPERATION”.
IfrebootingthedeviceandthetabletPCdonotcleartheproblems,pleasecontactthemanufacturer.
Errormessageexamples
[Warningmessage]
[Screenofnormaloperation]