8.2.Regularmaintenance(WeeklyChecklist)
Checkthattheboltsontheactuatorcoverarenotloose.
Checkthattheboltsonthebatterycoverarenotloose.
Checkthattheboltsonthethighsupportframearenotloose.
8.3.Cleaning
Riskofelectricalshockhazardduetomoisture!
Electricalshockmay be deliveredtothepatient and the userduetoexcessivemoisturewhen the device is
turnedonwhilecleaning.
‐ Donotcleanwhileinuse.
‐ Besuretoturnoffthepowerbeforecleaning.
Propertydamagemayoccurduetoimpropercleaningproceduresoragents!
Theproductmaybedamagedifitiscleanedanddisinfectedwithimproperproceduresoragents.
‐ Useonlythisprocedureforcleaning.
Ifthedeviceisusedbymorethanonepatient,itisessentialtocleanthedevicewithdisinfectantbeforeuse.
Topreventcross‐infectionbetweenpatients,thedevice(i.e.,waist part, thigh support frames, and thigh support
straps)shouldbecleanedaftereachuse.Thedeviceshouldbedisinfectedasfollows:
① Dampen(butdonotsoak)alint‐freeclothwithpurifiedwater.
② Wipethesurfaceofthedeviceforatleastfive(5)minutes.
③ Wipethesurfaceofthedevicewithan(isopropyl)alcohol‐baseddisinfectantforatleastfive(5)minutes.
‐RecommendedDisinfectant:Cavi‐Wipes(MetrexResearch,Kerr)orequivalent