Use Instructions

SAIT‐IFU‐GEMS‐HRev.1.1(2019‐11) Page62of79
8.2.Regularmaintenance(WeeklyChecklist)
Checkthattheboltsontheactuatorcoverarenotloose.
Checkthattheboltsonthebatterycoverarenotloose.
Checkthattheboltsonthethighsupportframearenotloose.
8.3.Cleaning
Riskofelectricalshockhazardduetomoisture!
Electricalshockmay be deliveredtothepatient and the userduetoexcessivemoisturewhen the device is
turnedonwhilecleaning.
Donotcleanwhileinuse.
Besuretoturnoffthepowerbeforecleaning.
Propertydamagemayoccurduetoimpropercleaningproceduresoragents!
Theproductmaybedamagedifitiscleanedanddisinfectedwithimproperproceduresoragents.
Useonlythisprocedureforcleaning.
Ifthedeviceisusedbymorethanonepatient,itisessentialtocleanthedevicewithdisinfectantbeforeuse.
Topreventcrossinfectionbetweenpatients,thedevice(i.e.,waist part, thigh support frames, and thigh support
straps)shouldbecleanedaftereachuse.Thedeviceshouldbedisinfectedasfollows:
Dampen(butdonotsoak)alint‐freeclothwithpurifiedwater.
Wipethesurfaceofthedeviceforatleastfive(5)minutes.
Wipethesurfaceofthedevicewithan(isopropyl)alcohol‐baseddisinfectantforatleastfive(5)minutes.
‐RecommendedDisinfectant:Cavi‐Wipes(MetrexResearch,Kerr)orequivalent