Chapter9.Troubleshooting
Propertydamagemayoccurduetoimproperrepair.
Ifpersonnelotherthanauthorizedbythemanufacturercarriesouttherepair,itmayirreversiblydamagethe
device.
Repairshouldonlybeperformedbythemanufacturer.
9.1.Troubleshooting
Issue Recommendedaction(s)
Thehipjointofthedevicebecomes
stuckduringoperation.
Iftheactuatorbecomesstuckduringoperation,donotapplyexcessive
forcetothedevice.
First,turnoffandthentakeoffthedevice.
Checkwhattheproblemiswhilegentlymovingthecorrespondingjoint.
Iftheproblempersistsoroccursrepeatedly,pleasecontactthe
manufacturer.
Case/Frame/Fastenerbreaks
Ifthehardwareisdamaged,donotoperatethedevice.Contactthe
manufacturer.
Thereisunusualnoisecomingfrom
thedevice.
Ifyouhearnoisefromthedevicethatisnotassociatedwithnormaluse,
pleasecontactthemanufacturer.
NoBluetoothconnection.
MakesurethatBluetoothisactivatedonthetabletPC.(Checkthisfrom
QuickSettingsonthetopofthetabletscreen).
RestarttheGEMS‐HApp.(Pressthebuttonleftofthehomebuttonon
thebottomofthetabletscreentocompletelyclosetheapp).
RestarttheGEMS‐HDevice.(Usethepowerswitchlocatedatthebottom
ofthecontrolpack).
RestarttheassociatedtabletPC.
Assistanceisuncomfortableorthe
patientcannotfeeltheassistive
force.
MakesuretheLEDindicatorisnotindicatinga“STOP”state.
RepeattheprocedureforresettingtheStandPosetozerowhenthe
patientisstationaryandstandingupright.Thisproceduremaybefound
insection6.8.3,under“Mitigationofrisksduetoexcessivejointangle
movement.”
Restarttheassistancealgorithmwithdefaultcontrolparametervalues
(orquittheGEMS‐Handrunitagain).
Checkwhetherthehipangle/torquevalueduringwalkingisnormalor
not.