Use Instructions
SAIT‐IFU‐GEMS‐HRev.1.1(2019‐11) Page75of79
Recommendedseparationdistancebetween
portableandmobileRFcommunicationsequipmentandtheGEMS‐H
TheGEMS‐HisintendedforuseinanelectromagneticenvironmentinwhichradiatedRFdisturbancesarecontrolled.
The customer or the user of the GEMS‐H can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the GEMS‐H as recommended below, according to the maximum
outputpowerofthecommunicationsequipment.
Ratedamaximum
outputpoweroftransmitter[W]
Separationdistanceaccordingtofrequencyoftransmitter[m]
IEC60601‐1‐2:2014
150kHzto80MHz
d=1.2
𝑝
80kHzto2.7GHz
d=2.0
𝑝
0.01 0.12 0.20
0.1 0.38 0.63
1 1.2 2.0
10 3.8 6.3
100 12 20
Fortransmittersratedamaximumoutputpowernotlistedabove,therecommendedseparationdistancedinmeters(m)canbeestimatedusing
theequationapplicabletothefrequencyofthetransmitter.whereisthepisthemaximumoutputpowerratingofthetransmitterinwatts(W)
accordingtothetransmittermanufacturer.
NOTE1:At80MHzand800MHz,theseparationdistanceforthehigherfrequencyrangeapplies
NOTE2:Theseguidelinesmaynotapplyinallsituations.Electromagneticpropagationisaffectedbyabsorptionandrefectionfromstructures,
objectsandpeople.
Guidanceandmanufacturer'sdeclaration‐electromagneticimmunity
TheGEMS‐Hisintendedforuseinanelectromagneticenvironment in which radiated RF disturbances are controlled. Portable RF
communications equipment should be used no closer than 30 cm (12 inches) to any part of the GEMS‐H. Otherwise, degradation of the
performanceofthisequipmentcouldresult.
Immunitytest Banda Service
a
Modulation
IEC60601
testlevel
Compliancelevel
Proximity fields
from
RFwireless
Communications
IEC61000‐4‐3
380‐390MHz TETRA400 Pulse modulation
18Hz
27V/m 27V/m
430–470MHz GMRS460
FRS460
FM
±5kHz
deviation
1kHzsine
28V/m 28V/m
704–787MHz LTEBand13,17 Pulse modulation
217Hz
9V/m 9V/m
800–960MHz GSM800:900
TETRA800
iDEN820
CDMA850
LTEBand5
Pulsemodulation
18Hz
28V/m 28V/m
1700–1990MHz GSM1800
CDMA1900
GSM1900
DECT
Pulse modulation
217Hz
28V/m 28V/m