User's Manual

11
Samsung LABGEO
PA20
Safety Warnings and Caution
Ŷ Cautions for system installation
Ŷ
Keep the analyzer at least 6 inches (15cm) away from any wall to provide
adequate ventilation and prevent overheating.
Ŷ
Be careful not to expose the analyzer to sources of vibration such as placing
it near a centrifuge.
Ŷ
Do not use the analyzer where liquids or chemicals are used or where gas
may be produced. Insure that no liquids are spilled into the Unit if there
is any spillage disconnect mains immediately and return to product for
serving.
Ŷ
The analyzer must be used at an operating temperature of 10 - 32 °C
(50- 89 °F). The analyzer might be aected by:
t Condensing humidity and water
t Heat and extreme temperature variations
t Electromagnetic radiation
t Movement of the analyzer during sample processing
Ŷ
Only use the cable provided with the analyzer.
Ŷ
To avoid risk of re or electric shock, do not use a damaged power cable.
Ŷ
Always use the correct voltage supply. Incorrect voltage can cause damage
to the analyzer or result in re. Be sure that the electrical outlet is properly
grounded.
Ŷ
Do not attempt to disassemble, repair or modify the analyzer as damage
could occur. Opening the unit will void the warranty. Contact Samsung
Technical Support.
Ŷ
Do not place any objects on the analyzer.
Ŷ
Do not cover the rear ventilation panel. This may damage the internal parts,
resulting in a re or electric shock.
Ŷ
Do not introduce foreign substances into the servicing hole.
Ŷ
Turn on the power switch at the rear of the Analyzer. (Grounding Proper
grounding is required when installing the system. Check the wall outlet
ground (Earth) for proper grounding to the facilities electrical ground. If you
are unsure of the outlet grounding, contact your facilities engineer to verify
the proper outlet ground). Next, turn the power switch on the front of the
Analyzer on and check that the program starts normally.
Ŷ
Do not touch the power plug with wet hands. This may result in an electric
shock.