Safety Warnings and Caution
Ŷ Cautions during use
Ŷ
This system is for professional use only. The user must be a physician, clinical
laboratory personnel, or healthcare professionals.
Ŷ
Avoid external shock or vibration during analysis.
Ŷ
Do not use expired discs.
Ŷ
Do not open the tray while the disc is being analyzed. During the analysis,
LED lamp or laser radiation may cause hazard due to improper operation.
Ŷ
The Analyzer contains UV/Visible LED Light source and laser. The users are
not exposed to Optical radiation in normal conditions. Because it is blocked
with analyzer cover cabinet. Do not open the tray during analysis. Do not
look into the analyzer during analysis. Your vision will be damaged due to
UV-lamp.
Ŷ
Powder-free gloves must be worn when operating the product.
Ŷ
Do not move the analyzer during operation.
Ŷ
When inserting a disc into the tray, check the orientation of the disc. It must
be inserted with the label side up. Consult test Instructions for Use.
Ŷ
Always use designated power cable. Do not unplug the device in the middle
of an analysis.
Ŷ
If the product produces smoke or strange odors, discontinue use, unplug
the power cord from the outlet, and contact your local distributor for service
information.
Ŷ
If operation has been aborted due to abnormal circumstances such as a
power outage during operation, discard the disc and start over with a new
disc.
Ŷ
Do not turn o the power during analysis. Press Cancel to abort a test if
needed. Discard the disc and start over with a new disc.
Ŷ
Be careful not to damage the barcode on the disc. If the barcode is
damaged, the test will not perform accurately.
Ŷ
Do not use the analyzer where liquids or chemicals are used or where gas
may be produced.
Ŷ
Do not allow liquid to spill into the analyzer. If there is any spillage,
disconnect power immediately and contact your local distributor.
Ŷ
If you wish to stop the analysis in progress without turning the analyzer o ,
press the Cancel button and stop the analysis. After stopping an analysis in
progress, throw away the cartridge and replace it with a new one.
Ŷ
The operating temperature of the analyzer is 10 - 32 °C (50 - 89 °F); the
analyzer automatically maintains an internal temperature of 37°C. (The
device shuts o automatically if it exceeds normal operating temperature.)